Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

, , , , , and

Published under licence by IOP Publishing Ltd
, , Citation Aitang Xing et al 2015 J. Phys.: Conf. Ser. 573 012054 DOI 10.1088/1742-6596/573/1/012054

1742-6596/573/1/012054

Abstract

3D Gamma index is one of the metrics which have been widely used for clinical routine patient specific quality assurance for IMRT, Tomotherapy and VMAT. The algorithms for calculating the 3D Gamma index using global and local methods implemented in two software tools: PTW- VeriSoft® as a part of OCTIVIUS 4D dosimeter systems and 3DVHTM from Sun Nuclear were assessed. The Gamma index calculated by the two systems was compared with manual calculated for one data set. The Gamma pass rate calculated by the two systems was compared using 3%/3mm, 2%/2mm, 3%/2mm and 2%/3mm for two additional data sets. The Gamma indexes calculated by the two systems were accurate, but Gamma pass rates calculated by the two software tools for same data set with the same dose threshold were different due to the different interpolation of raw dose data by the two systems and different implementation of Gamma index calculation and other modules in the two software tools. The mean difference was -1.3%±3.38 (1SD) with a maximum difference of 11.7%.