J D Wilson et al 2009 Physiol. Meas. 30 N1 doi:10.1088/0967-3334/30/1/N01
J D Wilson1, A J Adams2, P Murphy3, H Eswaran3 and H Preissl3,4
Show affiliationsThe design, safety analysis and performance of a fetal visual stimulation system suitable for fetal and neonatal magnetoencephalography studies are presented. The issue of fetal, neonatal and maternal safety is considered and the maximum permissible exposure is computed for the maternal skin and the adult eye. The risk for neonatal eye exposure is examined. It is demonstrated that the fetus, neonate and mother are not at risk.
87.80.-y Biophysical techniques (research methods)
85.25.Dq Superconducting quantum interference devices (SQUIDs)
87.50.C- Static and low-frequency electric and magnetic fields effects
Instrumentation and measurement
Issue 1 (January 2009)
Received 12 August 2008, accepted for publication 24 November 2008
Published 22 December 2008
J D Wilson et al 2009 Physiol. Meas. 30 N1