Vibration syndrome diagnosis using a cooling test verified by computerized photoplethysmography

Andrzej Dyszkiewicz1,2,3 and Michał Tendera4

Show affiliations


This study addresses the problem of vibration syndrome diagnosis by means of a cooling test verified by photoplethysmography. Measurement was taken on a small area on the fingertip plexus in which many arterio-venous anastomoses are present. In the opinion of many authors, flow disorders in this area are more typical for developing vibration syndrome than changes in the micro vessels. The study group comprised 128 subjects (58 women aged 40.9 ± 5.4 years and 70 men aged 38.7 ± 8.8 years) exposed to vibration. The control group consisted of 41 people (20 women aged 39.6 ± 7.3 years and 21 men aged 39.3 ± 6.4 years) who were not exposed to vibration. The patients were examined by a questionnaire and then a vibration perception threshold test and a cooling test were performed. The cooling test was verified both visually and using the computer method. Measurement data (S1, S2 and A) for each patient were obtained from averaging three pulse graphs. We departed from an average of 60 graphs (and more), the standard established in the literature, because of the cooling test specification, which causes huge thermodynamic parameter changeability in the plexus mass of the small finger under pulse waves coming one after another. A longer measurement time will reflect the thermal drift of the tested area in a direction to compensate for the reduced temperature. In the control group, all subjects showed an increase in planimetric indicators during the cooling test verified by computerized photoplethysmography. In the study group visual verification of the cooling test was positive in eight cases (6.2%) and the vibration perception threshold test was positive in seven cases (5.5%), but in computerized photoplethysmography the planimetric indicators decreased after cooling in 87 (67.4%) cases. Computer photoplethysmography is highly specific and shows greater sensitivity in detecting preclinical forms of vascular-type vibration syndrome when compared with palesthesiometry, the visually verified cooling test and the questionnaire. The proposed test enables the detection of vascular disorders in the prodromal period and gives time for preventive measures to be taken.


PACS

87.63.L- Visual imaging

02.70.Rr General statistical methods

47.63.Jd Microcirculation and flow through tissues

87.80.-y Biophysical techniques (research methods)

Subjects

Fluid dynamics

Computational physics

Instrumentation and measurement

Biological physics

Medical physics

Dates

Issue 4 (April 2006)

Received 2 August 2005, accepted for publication 17 January 2006

Published 16 February 2006



  1. Vibration syndrome diagnosis using a cooling test verified by computerized photoplethysmography

    Andrzej Dyszkiewicz and Michał Tendera 2006 Physiol. Meas. 27 353

  2. Teaching motion with the Global Positioning System

    Marko Budisa and Gorazd Planinsic 2003 Phys. Educ. 38 512

  3. Stochastic resonance

    Thomas Wellens et al 2004 Rep. Prog. Phys. 67 45

  4. The role of the rigged Hilbert space in quantum mechanics

    Rafael de la Madrid 2005 Eur. J. Phys. 26 287

  5. Designing meaningful density functional theory calculations in materials science—a primer

    Ann E Mattsson et al 2005 Modelling Simul. Mater. Sci. Eng. 13 R1

  6. Spectral Energy Distributions of Hard X-Ray Selected Active Galactic Nuclei in the XMM-Newton Medium Deep Survey

    M. Polletta et al. 2007 ApJ 663 81

  7. Self-similar Bianchi models: II. Class B models

    Pantelis S Apostolopoulos 2005 Class. Quantum Grav. 22 323

  8. Frequency- and space-resolved measurement of local density fluctuations in air by laser vibrometry

    B Hampel and J Woisetschläger 2006 Meas. Sci. Technol. 17 2835

  9. The Japanese space gravitational wave antenna—DECIGO

    Seiji Kawamura et al 2006 Class. Quantum Grav. 23 S125

  10. Distinguishability for magnetic resonance–electrical impedance tomography (MR–EIT)

    Haluk Altunel et al 2007 Phys. Med. Biol. 52 375

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.