Methods for the calibration of vibration pick-ups by laser interferometry: II. Experimental verification

, and

Published under licence by IOP Publishing Ltd
, , Citation Marek Dobosz et al 1998 Meas. Sci. Technol. 9 240 DOI 10.1088/0957-0233/9/2/011

0957-0233/9/2/240

Abstract

The regression analysis method applied for data transforming and error evaluation in the sine-approximation method (SAM) for the calibration of vibration pick-ups is experimentally verified. The new algorithm used for phase unwrapping in the SAM has also been examined. Detailed procedures for the analysis of data obtained from the SAM, the modified fringe counting method (FCM) and the minimum point method (MPM) are presented and a comparison of measurements taken has been carried out. The uncertainty of the FCM, suppressed by implementing fringe phase analysis in a single vibration cycle, is evaluated experimentally. The influence of distortions of the modulation phase signal is shown. The error budgets for all applied methods are given and compared. Expanded measurement uncertainties of about 0.2% for acceleration measurements and about 0.5% for sensitivity measurements (with a 99% confidence level) have been obtained.

Export citation and abstract BibTeX RIS