S Beguš and D Fefer 2007 Meas. Sci. Technol. 18 901 doi:10.1088/0957-0233/18/3/045
S Beguš and D Fefer
Show affiliationsA highly sensitive absorption-type nuclear magnetic resonance (NMR) magnetometer capable of measuring low magnetic fields below 1 mT with a NMR sample volume of 3 cm3 is described. It will be used for calibration purposes in an accredited calibration laboratory. The magnetometer uses a direct digital synthesizer for generation of the radio frequency signal, a sampling detector for demodulation of the radio frequency signal and a digital signal processor with an analogue-to-digital converter and a digital-to-analogue converter for signal processing using a lock-in technique with a digital phase sensitive detector. Frequency traceability of the NMR magnetometer reference oscillator is established by means of the long-wave transmitter DCF77.
07.55.Ge Magnetometers for magnetic field measurements
Issue 3 (March 2007)
Received 20 October 2006, in final form 17 January 2007
Published 13 February 2007
S Beguš and D Fefer 2007 Meas. Sci. Technol. 18 901