Determination of percentage depth dose for 6 and 10 MV x-rays using Ge-doped optical fibre and TLD-100

, and

Published 2 December 2002 Published under licence by IOP Publishing Ltd
, , Citation Y A Abdulla et al 2002 J. Radiol. Prot. 22 417 DOI 10.1088/0952-4746/22/4/306

0952-4746/22/4/417

Abstract

Percentage depth doses for 6 and 10 MV x-ray beams from a linear accelerator were measured using ∼1 cm long (∼0.3 mg) Ge-doped optical fibre as a thermoluminescence dosimeter for two field sizes, 5 × 5 and 10 × 10 cm2. The results indicate that the Ge-doped optical fibre dosimeter is in good agreement with the results from a PTW 30001 cylindrical ionisation chamber and TLD-100. For 6 MV x-ray beams we observe that the depth of maximum dose dmax is 1.5 and 2 cm for field sizes of 5 × 5 and 10 × 10 cm2 respectively. For 10 MV dmax is 2 cm for a field size of 5 × 5 cm2 and 2.5 cm for a 10 × 10 cm2 field.