Scan equalization digital radiography (SEDR) implemented with an amorphous selenium flat-panel detector: initial experience

Xinming Liu, Chao-Jen Lai, Lingyun Chen, Tao Han, Yuncheng Zhong, Youtao Shen, Tianpeng Wang and Chris C Shaw

Show affiliations


It is well recognized in projection radiography that low-contrast detectability suffers in heavily attenuating regions due to excessively low x-ray fluence to the image receptor and higher noise levels. Exposure equalization can improve image quality by increasing the x-ray exposure to heavily attenuating regions, resulting in a more uniform distribution of exposure to the detector. Image quality is also expected to be improved by using the slot-scan geometry to reject scattered radiation effectively without degrading primary x-rays. This paper describes the design of a prototype scan equalization digital radiography (SEDR) system implemented with an amorphous silicon (a-Si) thin-film transistor (TFT) array-based flat-panel detector. With this system, slot-scan geometry with alternate line erasure and readout (ALER) technique was used to achieve scatter rejection. A seven-segment beam height modulator assembly was mounted onto the fore collimator to regulate exposure regionally for chest radiography. The beam modulator assembly, consisting of micro linear motors, lead screw cartridge with lead (Pb) beam blockers attached, position feedback sensors and motor driver circuitry, has been tested and found to have an acceptable response for exposure equalization in chest radiography. An anthropomorphic chest phantom was imaged in the posterior–anterior (PA) view under clinical conditions. Scatter component, primary x-rays, scatter-to-primary ratios (SPRs) and primary signal-to-noise ratios (PSNRs) were measured in the SEDR images to evaluate the rejection and redistribution of scattered radiation, and compared with those for conventional full-field imaging with and without anti-scatter grid methods. SPR reduction ratios (SPRRRs, defined as the differences between the non-grid full-field SPRs and the reduced SPRs divided by the former) yielded approximately 59% for the full-field imaging with grid and 82% for the SEDR technique in the lungs, and 77% for the full-field imaging with grid and 95% for the SEDR technique in the subdiaphragm. The SEDR technique demonstrated a substantial improvement in PSNRs over the anti-scatter grid technique. The improvements of PSNRs varied with the regions and are more pronounced in heavily attenuating regions.


PACS

87.59.bf Digital radiography

87.56.Da Ancillary equipment

87.57.C- Image quality

Subjects

Medical physics

Dates

Issue 22 (21 November 2009)

Received 30 April 2009, in final form 13 October 2009

Published 4 November 2009



Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. Why do commercial CT scanners still employ traditional, filtered back-projection for image reconstruction?

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.