Depth of interaction calibration for PET detectors with dual-ended readout by PSAPDs

Yongfeng Yang1, Jinyi Qi1, Yibao Wu1, Sara St James1, Richard Farrell2, Purushottam A Dokhale2, Kanai S Shah2 and Simon R Cherry1

Show affiliations


Many laboratories develop depth-encoding detectors to improve the trade-off between spatial resolution and sensitivity in positron emission tomography (PET) scanners. One challenge in implementing these detectors is the need to calibrate the depth of interaction (DOI) response for the large numbers of detector elements in a scanner. In this work, we evaluate two different methods, a linear detector calibration and a linear crystal calibration, for determining DOI calibration parameters. Both methods can use measurements from any source distribution and location, or even the intrinsic lutetium oxyorthosilicate (LSO) background activity, and are therefore well suited for use in a depth-encoding PET scanner. The methods were evaluated by measuring detector and crystal DOI responses for all eight detectors in a prototype depth-encoding PET scanner. The detectors utilize dual-ended readout of LSO scintillator arrays with position-sensitive avalanche photodiodes (PSAPDs). The LSO arrays have 7 × 7 elements, with a crystal size of 0.92 × 0.92 × 20 mm3 and pitch of 1.0 mm. The arrays are read out by two 8 × 8 mm2 area PSAPDs placed at opposite ends of the arrays. DOI is measured by the ratio of the amplitude of the total energy signals measured by the two PSAPDs. Small variations were observed in the DOI responses of different crystals within an array as well as DOI responses for different arrays. A slightly nonlinear dependence of the DOI ratio on depth was observed and the nonlinearity was larger for the corner and edge crystals. The DOI calibration parameters were obtained from the DOI responses measured in a singles mode. The average error between the calibrated DOI and the known DOI was 0.8 mm if a linear detector DOI calibration was used and 0.5 mm if a linear crystal DOI calibration was used. A line source phantom and a hot rod phantom were scanned on the prototype PET scanner. DOI measurement significantly improved the image spatial resolution no matter which DOI calibration method was used. A linear crystal DOI calibration provided slightly better image spatial resolution compared with a linear detector DOI calibration.


PACS

87.57.uk Positron emission tomography (PET)

87.57.C- Image quality

87.63.-d Non-ionizing radiation equipment and techniques

87.57.U- Nuclear medicine imaging

Subjects

Medical physics

Dates

Issue 2 (21 January 2009)

Received 8 September 2008, in final form 24 November 2008

Published 19 December 2008



Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. SPECT detectors: the Anger Camera and beyond
  2. Review and current status of SPECT scatter correction
  3. Dynamic single photon emission computed tomography—basic principles and cardiac applications
More

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.