S Murty Goddu et al 2009 Phys. Med. Biol. 54 5663 doi:10.1088/0031-9155/54/19/001
S Murty Goddu1, Sasa Mutic, Olga L Pechenaya, Summer R Chaudhari, Jose Garcia-Ramirez, Dharanipathy Rangaraj, Eric E Klein, Deshan Yang, James Grigsby and Daniel A Low
Show affiliationsTomotherapy is an image-guided, intensity-modulated radiation therapy system that delivers highly conformal dose distributions in a helical fashion. This system is also capable of acquiring megavoltage computed-tomography images and registering them to the planning kVCT images for accurate target localization. Quality assurance (QA) of this device is time intensive, but can be expedited by improved QA tools and procedures. A custom-designed phantom was fabricated to improve the efficiency of daily QA of our Tomotherapy machine. The phantom incorporates ionization chamber measurement points, plugs of different densities and slide-out film cartridges. The QA procedure was designed to verify in less than 30 min the vital components of the tomotherapy system: static beam quality and output, image quality, correctness of image registration and energy of the helical dose delivery. Machine output, percent depth dose and off-axis factors are simultaneously evaluated using a static 5 × 40 cm2 open field. A single phantom scan is used to evaluate image quality and registration accuracy. The phantom can also be used for patient plan-specific QA. The QA results over a period of 6 months are reported in this paper. The QA process was found to be simple, efficient and capable of simultaneously verifying several important parameters.
For more information on this article see medicalphysicsweb.org
Issue 19 (7 October 2009)
Received 12 February 2009, in final form 22 May 2009
Published 1 September 2009
S Murty Goddu et al 2009 Phys. Med. Biol. 54 5663
J T Mendonça et al 2001 Meas. Sci. Technol. 12 1801
A Pakou et al 2005 J. Phys. G: Nucl. Part. Phys. 31 S1723
Alan A Coley and Sigbjørn Hervik 2004 Class. Quantum Grav. 21 5759
M Wieser et al 2005 Meas. Sci. Technol. 16 1667
B de Wit and H Nicolai 2001 Class. Quantum Grav. 18 3095
J A Vallejo 2001 J. Phys. A: Math. Gen. 34 9753
S D Mahanti and D Bilc 2004 J. Phys.: Condens. Matter 16 S5277
Merced Montesinos 2001 Class. Quantum Grav. 18 1847
Souichi Telada et al 2005 Class. Quantum Grav. 22 S975