Evaluation of the combined effects of target size, respiratory motion and background activity on 3D and 4D PET/CT images

Sang-June Park1, Dan Ionascu1, Joseph Killoran1, Marcelo Mamede2, Victor H Gerbaudo2, Lee Chin1 and Ross Berbeco1

Show affiliations


Gated (4D) PET/CT has the potential to greatly improve the accuracy of radiotherapy at treatment sites where internal organ motion is significant. However, the best methodology for applying 4D-PET/CT to target definition is not currently well established. With the goal of better understanding how to best apply 4D information to radiotherapy, initial studies were performed to investigate the effect of target size, respiratory motion and target-to-background activity concentration ratio (TBR) on 3D (ungated) and 4D PET images. Using a PET/CT scanner with 4D or gating capability, a full 3D-PET scan corrected with a 3D attenuation map from 3D-CT scan and a respiratory gated (4D) PET scan corrected with corresponding attenuation maps from 4D-CT were performed by imaging spherical targets (0.5–26.5 mL) filled with 18F-FDG in a dynamic thorax phantom and NEMA IEC body phantom at different TBRs (infinite, 8 and 4). To simulate respiratory motion, the phantoms were driven sinusoidally in the superior–inferior direction with amplitudes of 0, 1 and 2 cm and a period of 4.5 s. Recovery coefficients were determined on PET images. In addition, gating methods using different numbers of gating bins (1–20 bins) were evaluated with image noise and temporal resolution. For evaluation, volume recovery coefficient, signal-to-noise ratio and contrast-to-noise ratio were calculated as a function of the number of gating bins. Moreover, the optimum thresholds which give accurate moving target volumes were obtained for 3D and 4D images. The partial volume effect and signal loss in the 3D-PET images due to the limited PET resolution and the respiratory motion, respectively were measured. The results show that signal loss depends on both the amplitude and pattern of respiratory motion. However, the 4D-PET successfully recovers most of the loss induced by the respiratory motion. The 5-bin gating method gives the best temporal resolution with acceptable image noise. The results based on the 4D scan protocols can be used to improve the accuracy of determining the gross tumor volume for tumors in the lung and abdomen.


PACS

87.59.bd Computed radiography

87.19.U- Hemodynamics

87.57.uk Positron emission tomography (PET)

87.56.Da Ancillary equipment

87.57.N- Image analysis

87.57.C- Image quality

Subjects

Biological physics

Medical physics

Dates

Issue 13 (7 July 2008)

Received 31 March 2008, in final form 19 May 2008

Published 18 June 2008



  1. Evaluation of the combined effects of target size, respiratory motion and background activity on 3D and 4D PET/CT images

    Sang-June Park et al 2008 Phys. Med. Biol. 53 3661

  2. The electronic nature of high temperature cuprate superconductors

    M R Norman and C Pépin 2003 Rep. Prog. Phys. 66 1547

  3. Twenty-five questions for string theorists

    P Binétruy et al 2006 J. Phys. G: Nucl. Part. Phys. 32 129

  4. Electronic and ionic transport properties and other physical aspects of perovskites

    J B Goodenough 2004 Rep. Prog. Phys. 67 1915

  5. Compact directed percolation with movable partial reflectors

    Ronald Dickman and Daniel ben-Avraham 2002 J. Phys. A: Math. Gen. 35 7983

  6. Lattice kinetics of diffusion-limited coalescence and annihilation with sources

    E Abad et al 2002 J. Phys. A: Math. Gen. 35 1483

  7. Determination of void arrangements in inverse opals by transmission electron microscopy

    Christopher F Blanford et al 2006 J. Phys.: Conf. Ser. 26 264

  8. A selfconsistent theory of localization

    R Abou-Chacra et al 1973 J. Phys. C: Solid State Phys. 6 1734

  9. An evaluation of epoxy resin phantom materials for megavoltage photon dosimetry

    M Allahverdi et al 1999 Phys. Med. Biol. 44 1125

  10. Review of Particle Physics

    W-M Yao et al 2006 J. Phys. G: Nucl. Part. Phys. 33 1

Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. Why do commercial CT scanners still employ traditional, filtered back-projection for image reconstruction?

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.