Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer

Ning Wen, Huaiqun Guan1, Rabih Hammoud, Deepak Pradhan, T Nurushev, Shidong Li and Benjamin Movsas

Show affiliations


With the increased use of cone beam CT (CBCT) for daily patient setup, the accumulated dose from CBCT may be significantly higher than that from simulation CT or portal imaging. The objective of this work is to measure the dose from daily pelvic scans with fixed technical settings and collimations. CBCT scans were acquired in half-fan mode using a half bowtie and x-rays were delivered in pulsed-fluoro mode. The skin doses for seven prostate patients were measured on an IRB-approved protocol. TLD capsules were placed on the patient's skin at the central axis of three beams: AP, left lateral (Lt Lat) and right lateral (Rt Lat). To avoid the ring artefacts centred in the prostate, the treatment couch was dropped 3 cm from the patient's tattoo (central axis). The measured AP skin doses ranged 3–6 cGy for 20–33 cm separation. The larger the patient size the less the AP skin dose. Lateral doses did not change much with patient size. The Lt Lat dose was ~4.0 cGy, which was ~40% higher than the Rt Lat dose of ~2.6 cGy. To verify this dose asymmetry, surface doses on an IMRT QA phantom (oval shaped, 30 cm × 20 cm) were measured at the same three sites using TLD capsules with 3 cm table-drop. The dose asymmetry was due to: (1) kV source rotation which always starts from the patient's Lt Lat and ends at Lt Lat. Gantry rotation gets much slower near the end of rotation but dose rate stays constant and (2) 370° scan rotation (10° scan overlap on the Lt Lat side). In vivo doses were measured inside a Rando pelvic heterogeneous phantom using TLDs. The left hip (femoral head and neck) received the highest doses of ~10–11 cGy while the right hip received ~6–7 cGy. The surface and in vivo doses were also measured for phantoms at the central-axis setup. The difference was less than ~12% to the table-drop setup.


PACS

87.55.T- Record and verify systems and applications

87.53.Bn Dosimetry/exposure assessment

87.59.bd Computed radiography

87.56.Da Ancillary equipment

87.56.J- Collimation

87.55.-x Treatment strategy

Subjects

Medical physics

Dates

Issue 8 (21 April 2007)

Received 17 August 2006, in final form 11 December 2006

Published 2 April 2007



  1. Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer

    Ning Wen et al 2007 Phys. Med. Biol. 52 2267

  2. The nature of the inverse skin effect

    I D Culverwell et al 1989 Plasma Phys. Control. Fusion 31 387

  3. Equilibrium onions?

    L. Ramos et al 2004 Europhys. Lett. 66 888

  4. Electron-impact excitation of neutral boron using the R-matrix with the pseudostates method

    C P Ballance et al 2007 J. Phys. B: At. Mol. Opt. Phys. 40 1131

  5. Electro-thermo-elastic formulation for the analysis of smart structures

    Sheikh N Ahmad et al 2006 Smart Mater. Struct. 15 401

  6. Coupled oscillators, entangled oscillators, and Lorentz-covariant harmonic oscillators

    Y S Kim and Marilyn E Noz 2005 J. Opt. B: Quantum Semiclass. Opt. 7 S458

  7. Open-system nonequilibrium Green's functions and quantum transport in the transient regime

    Irena Knezevic and David K Ferry 2004 Semicond. Sci. Technol. 19 S220

  8. Optimal leaf sequencing with elimination of tongue-and-groove underdosage

    Srijit Kamath et al 2004 Phys. Med. Biol. 49 N7

  9. Crystallographic defects in CoCrPt thin film media: effect on interactions and magnetic viscosity

    H Laidler et al 2002 J. Phys. D: Appl. Phys. 35 512

  10. Characterization of the mechanical behavior of an electrically actuated microbeam

    Eihab M Abdel-Rahman et al 2002 J. Micromech. Microeng. 12 759

Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. The history of IRPA—up to the millennium
  2. Malignant pleural mesothelioma risk among nuclear workers: a review

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.