J J Gordon and J V Siebers 2007 Phys. Med. Biol. 52 1967 doi:10.1088/0031-9155/52/7/013
J J Gordon and J V Siebers
Show affiliationsThe van Herk margin formula (VHMF) relies on the accuracy of the convolution method (CM) to determine clinical target volume (CTV) to planning target volume (PTV) margins. This work (1) evaluates the accuracy of the CM and VHMF as a function of the number of fractions N and other parameters, and (2) proposes an alternative margin algorithm which ensures target coverage for a wider range of parameter values. Dose coverage was evaluated for a spherical target with uniform margin, using the same simplified dose model and CTV coverage criterion as were used in development of the VHMF. Systematic and random setup errors were assumed to be normally distributed with standard deviations Σ and σ. For clinically relevant combinations of σ, Σ and N, margins were determined by requiring that 90% of treatment course simulations have a CTV minimum dose greater than or equal to the static PTV minimum dose. Simulation results were compared with the VHMF and the alternative margin algorithm. The CM and VHMF were found to be accurate for parameter values satisfying the approximate criterion: σ[1 − γN/25] < 0.2, where γ = Σ/σ. They were found to be inaccurate for σ[1 − γN/25] > 0.2, because they failed to account for the non-negligible dose variability associated with random setup errors. These criteria are applicable when σ
σP, where σP = 0.32 cm is the standard deviation of the normal dose penumbra. (Qualitative behaviour of the CM and VHMF will remain the same, though the criteria might vary if σP takes values other than 0.32 cm.) When σ
σP, dose variability due to random setup errors becomes negligible, and the CM and VHMF are valid regardless of the values of Σ and N. When σ
σP, consistent with the above criteria, it was found that the VHMF can underestimate margins for large σ, small Σ and small N. A potential consequence of this underestimate is that the CTV minimum dose can fall below its planned value in more than the prescribed 10% of treatments. The proposed alternative margin algorithm provides better margin estimates and CTV coverage over the parameter ranges examined here. This algorithm is not amenable to expression as a simple formula (e.g., as a linear combination of Σ and σ). However, it can be easily calculated. For 0.1 cm ≤ σ ≤ 0.75 cm, 0 ≤ γ ≤ 1 and 5 ≤ N ≤ 30, the VHMF underestimates margins by as much as 33%. With the alternative margin algorithm, the maximum underestimate is 7%. These results suggest that the VHMF should be used with caution for hypofractionated treatment and in adaptive therapy.
Issue 7 (7 April 2007)
Received 19 September 2006, in final form 8 January 2007
Published 20 March 2007
J J Gordon and J V Siebers 2007 Phys. Med. Biol. 52 1967
Panagiota Kanti 2009 J. Phys.: Conf. Ser. 189 012020
Nigel G Adams et al 2009 J. Phys.: Conf. Ser. 192 012004
G T Andrews and M J Clouter 2001 Semicond. Sci. Technol. 16 679
Akira Yoshiasa et al 2009 J. Phys.: Conf. Ser. 190 012119
Tatsuma Nishioka 2009 Class. Quantum Grav. 26 242001
S M Bilenky et al 2008 J. Phys. G: Nucl. Part. Phys. 35 095003
U Kortshagen et al 1994 J. Phys. D: Appl. Phys. 27 301
F Paraguay-Delgado et al 2005 Nanotechnology 16 688
M Harfouche et al 2009 J. Phys.: Conf. Ser. 190 012189