Quick search Find article
Quick search
Find article

Practical and dosimetric implications of a new type of packaging for radiographic film

S Gillis and C De Wagter

Show affiliations


NOTE

Recently, Kodak introduced new light-tight packages (vacuum packaging, aluminium layer under black polyethylene and different paper) for their oncology films (EDR-2, X-Omat V and PPL-2). In order to avoid additional uncertainty and to ensure transferability of previously published results, we assessed in this study the effect of the old and new packages on the dosimetric response of EDR-2 radiographic film. Therefore, sensitometric measurements were performed for different film assemblies (new envelope + new paper, old envelope + old paper, new envelope without paper and old envelope without paper). In addition, to assess possible effects of the package on the film depth–dose response, packaged films were irradiated in parallel geometry, and central depth–dose curves were retrieved. For the perpendicular geometry, on the other hand, the effect of the package was assessed at large depth for a high intensity-modulated inverse-pyramid beam. The results of the sensitometric measurements reveal no difference between the packages. However, the white colour of the paper in both the packages induces a dose-dependent increase in optical density (0–0.12) of the film. The depth–dose curves show better reproducibility for the new package and the new paper improves the accuracy of film dosimetry, but despite the company's effort to evacuate the air out of the new envelope, it remains necessary to clamp the films in the phantom for the parallel irradiation geometry. At 5 cm depth, the films irradiated in parallel geometry show an under-response of 3–5% compared to films irradiated perpendicularly. Finally, even at locations of large photon scatter, no filtration effect from the aluminium layer incorporated in the new envelope has been observed for perpendicular irradiation geometry.


PACS

87.56.Da Ancillary equipment

87.55.T- Record and verify systems and applications

87.53.Bn Dosimetry/exposure assessment

Subjects

Medical physics

Dates

Issue 8 (21 April 2005)

Received 17 December 2004, in final form 14 February 2005

Published 30 March 2005



  1. Practical and dosimetric implications of a new type of packaging for radiographic film

    S Gillis and C De Wagter 2005 Phys. Med. Biol. 50 N63

  2. Comparison of He-Ne Lasers from the LMM and the IMGC Stabilized on 127I2 at 633 nm

    J de Vicente et al 1994 Metrologia 30 503

  3. Drag-free performance in a LISA mission with spherical proof masses

    B Lange 2002 Class. Quantum Grav. 19 1739

  4. Black hole evaporation without information loss

    C R Stephens et al 1994 Class. Quantum Grav. 11 621

  5. Magnetic order in HoNiBC and ErNiBC

    L J Chang et al 1996 J. Phys.: Condens. Matter 8 2119

  6. Unitarization of complex symmetric matrices

    G J N Brown and D S F Crothers 1994 J. Phys. A: Math. Gen. 27 2923

  7. A path integral derivation of χy-genus

    Guowu Meng 2003 J. Phys. A: Math. Gen. 36 1083

  8. The status of VIRGO

    F Acernese et al 2006 Class. Quantum Grav. 23 S63

  9. Integrated microfluidic systems for cell lysis, mixing/pumping and DNA amplification

    Chia-Yen Lee et al 2005 J. Micromech. Microeng. 15 1215

  10. Ionization processes in helium at pressures near to atmospheric

    J Dutton and D B Rees 1967 Br. J. Appl. Phys. 18 309

Users also read

What's this?
This innovative new feature generates a list of articles 'also read' by other users based on them reading the original article. Article abstracts citations and references are all considered and weighted accordingly. We hope that this will help you find relevant papers for your research.

  1. Variable transformation of calibration equations for radiation dosimetry
  2. The Schwarzschild effect of the dosimetry film Kodak EDR 2

Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. The concept and challenges of TomoTherapy accelerators

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.