Localization of sentinel nodes in breast cancer: novel method and device to help pen marking of active nodes during gamma camera imaging

Mikko S Laasanen1, Jari O Heikkinen1, Simo Saarakkala1 and Hannu Paajanen2

Show affiliations


NOTE

Gamma camera imaging with Tc-99m marking is a widely used method to locate sentinel lymph nodes (SNs) in breast cancer patients. Prior to SN biopsy, the anterior and lateral location of the SN is marked on the patient's skin using an ink pen. The pen marks guide the surgeon during an operation. However, in many cases the marking is difficult due to limited space under the detectors of a gamma camera. The aim of this study was to improve the pen marking method. Eleven female patients were imaged 3–4 h after injection of Tc-99m labelled Nanocol©. Injection was performed to parenchyma surrounding the breast tumour. To facilitate pen marking, two polycarbonate (PC) plates with 40 × 32 holes (spacing = 10 mm) were engineered for anterior and lateral side imaging and then installed on the bed of a dual-head gamma camera. Two drops of Tc-99m were placed into the top corners of both the PC plates, in order to trace the corresponding xy coordinates first from the acquired images and then from the plates. After imaging, the xy coordinates of the SN(s) were determined from the anterior and lateral side images. Subsequently, the location of each SN was marked with an ink pen on the skin through the small holes in the PC plates. According to the surgeon's evaluation, the distance between the marks and the true location of the SNs was 4.5 ± 6.9 mm. Measurements with a custom made phantom revealed that the accuracy of the novel method was significantly (P = 0.06) higher as compared with the traditional method (2.7 ± 3.0 mm versus 9.2 ± 3.0 mm). In addition, we were not able to mark the weakest activity (0.02 MBq) with the traditional method. Taken together, the marking process was considerably easier with the novel method, it had better accuracy and sensitivity than the traditional method and the device is simple enough to be adapted for most gamma cameras.


PACS

87.63.-d Non-ionizing radiation equipment and techniques

87.19.X- Diseases

07.85.Fv X- and gamma-ray sources, mirrors, gratings, and detectors

Subjects

Instrumentation and measurement

Biological physics

Medical physics

Dates

Issue 7 (7 April 2005)

Received 10 November 2004, in final form 7 January 2005

Published 16 March 2005



  1. Localization of sentinel nodes in breast cancer: novel method and device to help pen marking of active nodes during gamma camera imaging

    Mikko S Laasanen et al 2005 Phys. Med. Biol. 50 N49

  2. Sideband structure in angle-resolved electron spectra from laser-assisted Auger decay generated by ultra-short pulses

    A K Kazansky and N M Kabachnik 2009 J. Phys. B: At. Mol. Opt. Phys. 42 121002

  3. Excited-state imaging of cold atoms

    David V Sheludko et al 2007 J. Phys.: Conf. Ser. 80 012040

  4. Ionization cross section ratios of rare-gas atoms (Ne, Ar, Kr and Xe) by electron impact from threshold to 1 keV

    Ataru Kobayashi et al 2002 J. Phys. B: At. Mol. Opt. Phys. 35 2087

  5. Possible variations of the fine structure constant α and their metrological significance

    K A Bronnikov and S A Kononogov 2006 Metrologia 43 R1

  6. An approximate analogue solution for the axially symmetrical form of the biharmonic equation

    A R Curtis and N P R Sherry 1977 J. Phys. D: Appl. Phys. 10 481

  7. Buckling of rectangular functionally graded material plates subjected to nonlinearly distributed in-plane edge loads

    X L Chen and K M Liew 2004 Smart Mater. Struct. 13 1430

  8. Comment on `A systematic review of the precision and accuracy of dose measurements in photon radiotherapy using polymer and Fricke MRI gel dosimetry'

    Y De Deene et al 2003 Phys. Med. Biol. 48 L15

  9. X-shape oligo(thiophene)s as donor materials for vacuum-deposited organic photovoltaic cells

    Wang Ya-Nan et al 2008 Chinese Phys. B 17 1448

  10. Improved metal cluster deposition on a genetically engineered tobacco mosaic virus template

    Sang-Yup Lee et al 2005 Nanotechnology 16 S435

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.