Electron beam treatment verification using measured and Monte Carlo predicted portal images

G Jarry and F Verhaegen

Show affiliations


Electron beam treatments may benefit from techniques to verify patient positioning and dose delivery. This is particularly so for complex techniques such as mixed photon and electron beam radiotherapy and electron beam modulated therapy. This study demonstrates that it is possible to use the bremsstrahlung photons in an electron beam from a dual scattering foil linear accelerator to obtain portal images of electron beam treatments. The possibility of using Monte Carlo (MC) simulations to predict the electron beam treatment portal images was explored. The MC code EGSnrc was used to model a Varian CL21EX linear accelerator (linac) and to characterize the bremsstrahlung photon production in the linac head. It was found that the main sources of photons in the electron beam are the scattering foils, the applicator and the beam-shaping cut-out. Images were acquired using the Varian CL21EX linac and the Varian aS500 electronic portal imager (EPI); four electron energies (6, 9, 12, 16 MeV), and different applicator and cut-out sizes were used. It was possible to acquire images with as little as 10.7 MU per image. The contrast, the contrast-to-noise ratio (CNR), the signal-to-noise ratio (SNR), the resolution and an estimate of the modulated transfer function (MTF) of the electron beam portal images were computed using a quality assurance (QA) phantom and were found to be comparable to those of a 6 MV photon beam. Images were also acquired using a Rando anthropomorphic phantom. MC simulations were used to model the aS500 EPID and to obtain predicted portal images of the QA and Rando phantom. The contrast in simulated and measured portal images agrees within ±5% for both the QA and the Rando phantom. The measured and simulated images allow for a verification of the phantom positioning by making sure that the structure edges are well aligned. This study suggests that the Varian aS500 portal imager can be used to obtain patient portal images of electron beams in the scattering foil linacs.


PACS

87.55.K- Monte Carlo methods

87.55.Qr Quality assurance in radiotherapy

87.55.km Verification

87.56.Da Ancillary equipment

87.53.Bn Dosimetry/exposure assessment

Subjects

Medical physics

Dates

Issue 21 (7 November 2005)

Received 26 May 2005, in final form 8 August 2005

Published 13 October 2005



Users also read

What's this?
This innovative new feature generates a list of articles 'also read' by other users based on them reading the original article. Article abstracts citations and references are all considered and weighted accordingly. We hope that this will help you find relevant papers for your research.

  1. Optimization of image quality and dose for Varian aS500 electronic portal imaging devices (EPIDs)
  2. Scanning tunnelling microscope imaging of nanoscale electron density gradients on the surface of GaAs
  3. The electronic structure of magnetic transition metallic materials
More

Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. The history of IRPA—up to the millennium
  2. Malignant pleural mesothelioma risk among nuclear workers: a review

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.