High-resolution PET detector design: modelling components of intrinsic spatial resolution

Jennifer R Stickel and Simon R Cherry

Show affiliations


The development of dedicated small animal PET (positron emission tomography) scanners has led to significantly higher spatial resolution and comparable sensitivity to clinical scanners. However, it is not clear whether we are approaching the fundamental limit of spatial resolution. This work aims to understand what is currently limiting spatial resolution during data formation and collection and how to apply that knowledge to obtain the best possible resolution for small animal PET without sacrificing sensitivity. Monte Carlo simulations were performed of the interactions of a 511 keV photon in a variety of detector materials to evaluate the modulation transfer function of the materials. Positron range, non-colinearity and pixel size were modelled to determine the contribution of additional components of data formation and collection on the complete modulation transfer function of a PET system. These simulations are shown to predict the intrinsic detector resolution of current high resolution systems very well. They also show that current detectors are not limited by inter-crystal scatter. An intrinsic resolution of 0.5 mm can be achieved, but would require a detector with a pixel size of around 250 µm that can be read out unambiguously. It is shown that a range of different detector materials, both scintillators and semiconductors, can be used in these high-resolution detectors. While this design relies on thin (~3 mm) pieces of material, stacks of the material are shown to simultaneously provide spatial resolution near 0.5 mm and 60% efficiency. This work has shown that detectors with significantly better resolution and sensitivity can be developed for small animal PET applications.


PACS

87.57.uk Positron emission tomography (PET)

02.50.Ng Distribution theory and Monte Carlo studies

87.63.-d Non-ionizing radiation equipment and techniques

87.57.U- Nuclear medicine imaging

87.57.C- Image quality

Subjects

Computational physics

Medical physics

Dates

Issue 2 (21 January 2005)

Received 19 September 2004, in final form 16 November 2004

Published 23 December 2004



  1. High-resolution PET detector design: modelling components of intrinsic spatial resolution

    Jennifer R Stickel and Simon R Cherry 2005 Phys. Med. Biol. 50 179

  2. Coherent control of autoionization in optically dense media

    David Petrosyan and P Lambropoulos 2001 J. Phys. B: At. Mol. Opt. Phys. 34 1711

  3. A distortion of pseudotetramers coupled with the Jahn–Teller effect in the geometrically frustrated spinel system CdV2O4

    Masashige Onoda and Junichi Hasegawa 2003 J. Phys.: Condens. Matter 15 L95

  4. Charge transfer in H+-Na0 collisions: atomic orbital calculations

    R Shingal et al 1986 J. Phys. B: At. Mol. Phys. 19 309

  5. Optimization of calcination conditions on the Bi-2223 kinetic formation and grain size

    V Garnier et al 2000 Supercond. Sci. Technol. 13 602

  6. Dynamics of vortices in quasi-two-dimensional classical Heisenberg magnets with weak easy-plane anisotropy

    A R Volkel et al 1992 J. Phys.: Condens. Matter 4 9411

  7. Reversors and symmetries for polynomial automorphisms of the complex plane

    A Gómez and J D Meiss 2004 Nonlinearity 17 975

  8. Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes

    Yongfeng Yang et al 2006 Phys. Med. Biol. 51 2131

  9. On the scattering method for the \overline{\partial} -equation and reconstruction of convection coefficients

    Alexandru Tamasan 2004 Inverse Problems 20 1807

  10. Falling faster than in free fall?

    J D Hey et al 2004 Eur. J. Phys. 25 63

Users also read

What's this?
This innovative new feature generates a list of articles 'also read' by other users based on them reading the original article. Article abstracts citations and references are all considered and weighted accordingly. We hope that this will help you find relevant papers for your research.

  1. Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution

Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. SPECT detectors: the Anger Camera and beyond
  2. Review and current status of SPECT scatter correction
  3. Dynamic single photon emission computed tomography—basic principles and cardiac applications
More

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.