Cardiac PET imaging in mice with simultaneous cardiac and respiratory gating

Yongfeng Yang1, Stephen Rendig1, Stefan Siegel2, Danny F Newport2 and Simon R Cherry1

Show affiliations


Gating firmware and software were developed for the microPET II small animal scanner. The measured cardiac and respiratory signals were collected and converted to TTL gating signals by a Biopac MP150 data acquisition system and sent to microPET II through two BNC connectors on the front panel. During acquisition, the coincidence monitor takes the average of the last eight gate input cycles and inserts this into the list mode data stream on the falling edge of the gating pulse. This value is then used to determine the current time interval of the next gate cycle when the list mode data are sorted into sinograms. The gating firmware and software were validated by an experiment using a rotating point source. Mouse heart (18F-FDG) and bone (18F) imaging was performed with simultaneous cardiac and respiratory gating. It was clearly demonstrated that the contractile function of the mouse heart can be studied by cardiac-gated imaging with microPET II. The left ventricular volumes at different times of the cardiac cycle were measured and the ejection fraction was calculated. In the bone scan, no detectable movement caused by heart contraction was observed. Respiratory motion was more subtle with virtually no motion for more than 75% of the respiratory cycle. The motion of the mouse heart and bones in the thorax caused by respiration was less than 1 mm. It appears with the current resolution of PET, and the small fraction of the respiratory cycle in which motion occurs, that respiratory gating is probably not necessary for most mouse cardiac studies.


PACS

87.57.uk Positron emission tomography (PET)

87.19.Hh Cardiac dynamics

47.63.Cb Blood flow in cardiovascular system

87.57.U- Nuclear medicine imaging

87.63.-d Non-ionizing radiation equipment and techniques

87.19.U- Hemodynamics

Subjects

Fluid dynamics

Biological physics

Medical physics

Dates

Issue 13 (7 July 2005)

Received 11 January 2005, in final form 13 April 2005

Published 8 June 2005



Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. SPECT detectors: the Anger Camera and beyond
  2. Review and current status of SPECT scatter correction
  3. Dynamic single photon emission computed tomography—basic principles and cardiac applications
More

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.