B J Pichler et al 2004 Phys. Med. Biol. 49 4305 doi:10.1088/0031-9155/49/18/008
B J Pichler1, B K Swann2, J Rochelle2, R E Nutt2, S R Cherry1 and S B Siegel2
Show affiliationsAvalanche photodiodes (APDs) have proven to be useful as light detectors for high resolution positron emission tomography (PET). Their compactness makes these devices excellent candidates for replacing bulky photomultiplier tubes (PMTs) in PET systems where space limitations are an issue. The readout of densely packed, 10 × 10 lutetium oxyorthosilicate (LSO) block detectors (crystal size 2.0 × 2.0 × 12 mm3) with custom-built monolithic 3 × 3 APD arrays was investigated. The APDs had a 5 × 5 mm2 active surface and were arranged on a 6.25 mm pitch. The dead space on the edges of the array was 1.25 mm. The APDs were operated at a bias voltage of approximately 380 V for a gain of 100 and a dark current of 10 nA per APD. The standard deviation in gain between the APDs in the array ranged from 1.8 to 6.5% as the gain was varied from 50 to 108. A fast, low-noise, multi-channel charge sensitive preamplifier application-specific integrated circuit (ASIC) was developed for the APD readout. The amplifier had a rise time of 8 ns, a noise floor of 515 e− rms and a 9 e− pF−1 noise slope. An acquired flood image showed that all 100 crystals from the block detector could be resolved. Timing measurements with single-channel LSO-APD detectors, as well as with the array, against a plastic scintillator and PMT assembly showed a time resolution of 1.2 ns and 2.5 ns, respectively. The energy resolution measured with a single 4.0 × 4.0 × 10 mm3 LSO crystal, wrapped in four-layer polytetrafluoroethylene (PTFE) tape and coupled with optical grease on a single APD of the array, yielded 15% (full width at half maximum, FWHM) at 511 keV. Stability tests over 9 months of operation showed that the APD arrays do not degrade appreciably. These results demonstrate the ability to decode densely packed LSO scintillation blocks with compact APD arrays. The good timing and energy resolution makes these detectors suitable for high resolution PET.
87.57.uk Positron emission tomography (PET)
87.63.-d Non-ionizing radiation equipment and techniques
Issue 18 (21 September 2004)
Received 17 April 2004
Published 3 September 2004
B J Pichler et al 2004 Phys. Med. Biol. 49 4305
David B Lobell and Christopher B Field 2007 Environ. Res. Lett. 2 014002
Michał Hanasz et al 2009 ApJ 706 L155
Aurora Sicilia-Aguilar et al. 2007 ApJ 659 1637
S.-B. Qian et al 2009 ApJ 706 L96
Hervé Cottin et al. 2003 ApJ 590 874
M. Brüggen et al 2005 ApJ 631 L21
E. Panko et al. 2009 The Astronomical Journal 138 1709
Changhui Li and Lihong V Wang 2009 Phys. Med. Biol. 54 R59
Ken B. Henisey et al. 2009 ApJ 706 705