E Pantelis et al 2004 Phys. Med. Biol. 49 55 doi:10.1088/0031-9155/49/1/004
E Pantelis1, P Papagiannis1, G Anagnostopoulos2, D Baltas2, P Karaiskos1,3, P Sandilos3,4 and L Sakelliou1
Show affiliationsA simple, time efficient, analytical model incorporating heterogeneities and body dimensions around a point 192Ir source is generalized for accurate dosimetry around commercially available 192Ir brachytherapy sources. The generalized model was verified in dosimetry of a clinical 192Ir high dose rate prostate monotherapy application, involving 16 catheters and 83 source dwell positions, through comparison with corresponding treatment planning system data. The computational time efficiency and accuracy of the proposed model allowed the assessment of the impact that uncertainties in source dwell positions and catheter reconstruction may have on dose distributions, and how these could potentially affect the clinical outcome. Results revealed that a 0.1 cm catheter reconstruction uncertainty and a 0.15 cm source position uncertainty along the catheter lead to a dose uncertainty of less than 2% for doses lower than 200% of the prescribed dose, reaching up to 5% for points lying in close proximity to the catheters. These uncertainties were found to have no impact (less than 1%) on dose volume histogram results of both the planning target volume and the urethra. A catheter reconstruction uncertainty as high as 0.2 cm results in a dose uncertainty greater than 2%, reaching up to 9%, only for points inside the 150% contour. However, even in this case, the impact on dose volume histogram calculations is less than 3%.
87.53.Jw Therapeutic applications, including brachytherapy
Issue 1 (7 January 2004)
Received 16 July 2003, in final form 8 October 2003
Published 15 December 2003
E Pantelis et al 2004 Phys. Med. Biol. 49 55
G M Cicuta et al 2002 J. Phys. A: Math. Gen. 35 1125
David Sowby 2008 J. Radiol. Prot. 28 271
2009 J. Radiol. Prot. 29
J Swenson et al 1996 J. Phys.: Condens. Matter 8 3545
E Blanco et al 2006 Plasma Phys. Control. Fusion 48 699
Young S Kim et al 2004 J. Opt. B: Quantum Semiclass. Opt. 6
K O Kvashnina et al 2009 J. Phys.: Conf. Ser. 190 012191
Karl K Irikura et al 2004 Metrologia 41 369
L Castellani et al 1990 Class. Quantum Grav. 7 1767