Quick search Find article
Quick search
Find article

Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation

Thomas Bortfeld, Kimmo Jokivarsi, Michael Goitein, Jong Kung and Steve B Jiang

Show affiliations


There has been some concern that organ motion, especially intra-fraction organ motion due to breathing, can negate the potential merit of intensity-modulated radiotherapy (IMRT). We wanted to find out whether this concern is justified. Specifically, we wanted to investigate whether IMRT delivery techniques with moving parts, e.g., with a multileaf collimator (MLC), are particularly sensitive to organ motion due to the interplay between organ motion and leaf motion. We also wanted to know if, and by how much, fractionation of the treatment can reduce the effects.

We performed a statistical analysis and calculated the expected dose values and dose variances for volume elements of organs that move during the delivery of the IMRT. We looked at the overall influence of organ motion during the course of a fractionated treatment. A linear-quadratic model was used to consider fractionation effects. Furthermore, we developed software to simulate motion effects for IMRT delivery with an MLC, with compensators, and with a scanning beam. For the simulation we assumed a sinusoidal motion in an isocentric plane.

We found that the expected dose value is independent of the treatment technique. It is just a weighted average over the path of motion of the dose distribution without motion. If the treatment is delivered in several fractions, the distribution of the dose around the expected value is close to a Gaussian. For a typical treatment with 30 fractions, the standard deviation is generally within 1% of the expected value for MLC delivery if one assumes a typical motion amplitude of 5 mm (1 cm peak to peak). The standard deviation is generally even smaller for the compensator but bigger for scanning beam delivery. For the latter it can be reduced through multiple deliveries ('paintings') of the same field.

In conclusion, the main effect of organ motion in IMRT is an averaging of the dose distribution without motion over the path of the motion. This is the same as for treatments with conventional beams. Additional effects that are specific to the IMRT delivery technique appear to be relatively small, except for the scanning beam.


PACS

87.56.ng Wedges and compensators

87.56.J- Collimation

87.55.-x Treatment strategy

87.55.Gh Simulation

87.19.U- Hemodynamics

Subjects

Biological physics

Medical physics

Dates

Issue 13 (7 July 2002)

Received 26 February 2002

Published 20 June 2002



  1. Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation

    Thomas Bortfeld et al 2002 Phys. Med. Biol. 47 2203

  2. High-throughput and data mining with ab initio methods

    Dane Morgan et al 2005 Meas. Sci. Technol. 16 296

  3. Measurements of inclusive W and Z cross sections in p\overline{p} collisions at \sqrt{s} = 1\hbox{.}96\, {{\rm TeV}}

    A Abulencia et al 2007 J. Phys. G: Nucl. Part. Phys. 34 2457

  4. Analysis of an exact inversion algorithm for spiral cone-beam CT

    Alexander Katsevich 2002 Phys. Med. Biol. 47 2583

  5. Local-density approximation for the exchange energy functional in excited-state density functional theory

    Prasanjit Samal and Manoj K Harbola 2005 J. Phys. B: At. Mol. Opt. Phys. 38 3765

  6. Joint penalized-likelihood reconstruction of time-activity curves and regions-of-interest from projection data in brain PET

    E Krestyannikov et al 2008 Phys. Med. Biol. 53 2877

  7. Unicyclic components in random graphs

    E Ben-Naim and P L Krapivsky 2004 J. Phys. A: Math. Gen. 37 L189

  8. Bulk optical properties of healthy female breast tissue

    T Durduran et al 2002 Phys. Med. Biol. 47 2847

  9. Scrutinizing Exotic Cosmological Models Using ESSENCE Supernova Data Combined with Other Cosmological Probes

    T. M. Davis et al. 2007 ApJ 666 716

  10. Spin dynamics in Ho2Ru2O7

    J S Gardner et al 2005 J. Phys.: Condens. Matter 17 7089

Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. The concept and challenges of TomoTherapy accelerators

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.