J B Van de Kamer et al 2001 Phys. Med. Biol. 46 183 doi:10.1088/0031-9155/46/1/313
J B Van de Kamer, J J W Lagendijk, A A C De Leeuw and H Kroeze
Show affiliations Present-day regional hyperthermia treatment planning systems are limited to centimetre resolution. To obtain CT-resolution SAR distributions, a method called quasistatic zooming has been developed: using the centimetre-resolution
-field distribution and the CT-resolution tomogram, the CT-resolution SAR distribution is obtained. For a low frequency of 10 MHz this method has been validated sucessfully using CT-resolution SAR computations.
It appears that these CT-resolution SAR distributions are completely different from centimetre-resolution SAR distributions, indicating the necessity for high-resolution SAR modelling. Using the presented zooming technique, reliable CT-resolution SAR modelling is now possible with relatively short computation times. So far, the zooming method has only been validated for low frequencies, but clinically relevant frequencies appear to be possible.
87.50.wp Therapeutic applications
87.19.Pp Biothermics and thermal processes in biology
Issue 1 (January 2001)
Received 21 August 2000
J B Van de Kamer et al 2001 Phys. Med. Biol. 46 183
A Bette and S Zakrzewski 1997 J. Phys. A: Math. Gen. 30 195
Róbert Horváth et al 2003 J. Micromech. Microeng. 13 419
V. L. Bulatov and P. E. Kornilovitch 2005 Europhys. Lett. 71 352
H Dimmelmeier et al 2002 Class. Quantum Grav. 19 1291
P Moskal et al 2003 J. Phys. G: Nucl. Part. Phys. 29 2235
Theodoros P Horikis 2009 J. Phys. A: Math. Theor. 42 442004
Simon Catterall and Toby Wiseman JHEP12(2007)104
B Protas 2008 Nonlinearity 21 R203
Xifeng Lu et al 2009 J. Phys.: Conf. Ser. 188 012010