Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation

K L Pasma, M Kroonwijk, J C J de Boer, A G Visser and B J M Heijmen

Show affiliations


Measuring portal dose with an electronic portal imaging device (EPID) in external beam radiotherapy can be used to perform routine dosimetric quality control checks on linear accelerators and to verify treatments ( in vivo dosimetry). An accurate method to measure portal dose images (PDIs) with a commercially available fluoroscopic EPID has been developed. The method accounts for (i) the optical `cross talk' within the EPID structure, (ii) the spatially non-uniform EPID response and (iii) the nonlinearity of the EPID response. The method is based on a deconvolution algorithm. Measurement of the required input data is straightforward. The observed nonlinearity of the EPID response was largely due to the somewhat outdated EPID electronics. Nonlinearity corrections for more modern systems are expected to be smaller. The accuracy of the method was assessed by comparing PDIs measured with the EPID with PDIs measured with a scanning ionization chamber in a miniphantom, located at the same position as the fluorescent screen. For irradiations in open, wedged and intensity modulated 25 MV photon beams (produced with dynamic multileaf collimation) EPID and ionization chamber measurements agreed to within 1% (1 SD).


PACS

87.56.ng Wedges and compensators

87.56.Da Ancillary equipment

87.56.Fc Quality assurance equipment

87.56.J- Collimation

87.55.km Verification

87.55.T- Record and verify systems and applications

Subjects

Medical physics

Dates

Issue 8 (August 1998)

Received 18 November 1997, in final form 25 February 1998



Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. The concept and challenges of TomoTherapy accelerators

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.