P R Hoskins et al 1989 Phys. Med. Biol. 34 1709 doi:10.1088/0031-9155/34/11/018
P R Hoskins, T Anderson and W N McDicken
Show affiliationsA flow phantom for the generation of physiological Doppler waveforms is described. The suspension of scattering particles is driven by a gear pump powered by a stepping motor. The speed of the stepping motor is controlled by a BBC microcomputer. The waveform shape is selected from a library of waveforms from disc. Use of the microcomputer allows the waveform shape and mean flow to be easily changed. Sephadex particles suspended in a solution of glycerol were used as artificial blood. Thin walled heat shrink tubing which had been moulded around metal rods was used. Distortions in the waveforms caused by reflections from the end of the tubing were largely removed by reducing the pipe diameter to half of its value for 30 cm from the end of the pipe. There was good agreement between the control waveforms and the Doppler waveforms over a wide range of waveform pulsatility.
43.80.Vj Acoustical medical instrumentation and measurement techniques
Issue 11 (November 1989)