SCIENTIFIC NOTE

Induced activity in a high-energy linear accelerator

and

Published under licence by IOP Publishing Ltd
, , Citation L Ahlgren and L E Olsson 1988 Phys. Med. Biol. 33 351 DOI 10.1088/0031-9155/33/3/004

0031-9155/33/3/351

Abstract

Activation products which contribute significantly to the exposure of the staff operating the accelerator were found in the accelerator head in the transparent holder for the lead blocks and in the 60 degrees wedge filter made of lead. A proper choice of material in these parts of the accelerator could easily reduce this exposure without excessive changes in the construction. The absorbed dose-rate level should always be checked before maintenance is carried out inside the accelerator head.