Comparison of the standards for air kerma of the BEV and the BIPM for 60Co gamma radiation

C Kessler1, P J Allisy-Roberts1, A Steurer2, W Tiefenboeck2 and F Gabris2

Show affiliations


KEY COMPARISON

A direct comparison of the standards for air kerma of the Bundesamt für Eich- und Vermessungswesen (BEV), Austria, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 60Co radiation beam of the BIPM in March 2009. The result, expressed as a ratio of the BEV and the BIPM standards for air kerma, is 1.0034 with a combined standard uncertainty of 2.1 × 10−3. The result of the previous direct comparison in 60Co γ rays, made in 1995, was 1.0055 (25).

The degrees of equivalence between the BEV and the other participants in the key comparison have been calculated and the results are presented in the form of a matrix. A graphical presentation is also given.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A (Technical Supplement 2009)



  1. Comparison of the standards for air kerma of the BEV and the BIPM for 60Co gamma radiation

    C Kessler et al 2010 Metrologia 47 06006

  2. Update of the ongoing comparison BIPM.RI(II)-K1.Co-60 including activity measurements of the radionuclide 60Co for the IFIN-HH (Romania), NIST (USA) and the BEV (Austria)

    C Michotte et al 2010 Metrologia 47 06010

  3. Update report of the BIPM comparison BIPM.RI(II)-K1.Na-22 of activity measurements of the radionuclide 22Na to include the CIEMAT, PTB and the NMISA

    C Michotte et al 2010 Metrologia 47 06001

  4. Comparison of the standards for air kerma of the ITN (Portugal) and the BIPM for 137Cs γ-rays

    C Kessler et al 2009 Metrologia 46 06012

  5. APMP comparison of the activity measurements of Ba-133 (APMP.RI(II)-K2.Ba-133) and links to the SIR

    A Yunoki et al 2009 Metrologia 46 06014

  6. Comparison of the standards for air kerma of the BEV and the BIPM for 137Cs gamma radiation

    C Kessler et al 2010 Metrologia 47 06007

  7. Comparison of the standards for air kerma of the NIM and the BIPM for 60Co gamma radiation

    P J Allisy-Roberts et al 2009 Metrologia 46 06013

  8. BIPM comparison BIPM.RI(II)-K1.Cu-64 of the activity measurements of the radionuclide 64Cu

    C Michotte et al 2009 Metrologia 46 06010

  9. Comparison of the standards for absorbed dose to water of the VNIIFTRI, Russia and the BIPM in 60Co γ rays

    P J Allisy-Roberts et al 2010 Metrologia 47 06003

  10. Comparison of the standards for air kerma of the NPL and the BIPM for 60Co γ rays

    C Kessler et al 2010 Metrologia 47 06004