S Solve et al 2010 Metrologia 47 01004 doi:10.1088/0026-1394/47/1A/01004
S Solve1, R Chayramy1, M Stock1, J Nicolas2 and A Van Theemsche2
Show affiliationsA comparison of the 10 V Josephson array voltage standard of the Bureau International des Poids et Mesures (BIPM) was made with that of the Service Métrologie – Metrologische Dienst (SMD), Belgium, in November 2009. For this exercise, the option B of the BIPM.EM-K10.b comparison protocol was applied. The results of both participants are in very good agreement and the overall relative standard uncertainty is 1.3 parts in 1010.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Issue 1A (Technical Supplement 2010)
S Solve et al 2010 Metrologia 47 01004
R Goebel et al 2010 Metrologia 47 01005
B M Wood and S Solve 2009 Metrologia 46 R13
J D Dow and C P Flynn 1980 J. Phys. C: Solid State Phys. 13 1341
P Suresh Kumar et al 2002 J. Opt. A: Pure Appl. Opt. 4 247
S Thomas Lee et al 2003 Meas. Sci. Technol. 14 858
Carl M Bender et al 2004 J. Phys. A: Math. Gen. 37 147
R Cech et al 2007 Phys. Med. Biol. 52 879
Aman Russom et al 2009 New J. Phys. 11 075025
R. Abbasi et al. 2009 ApJ 701 1721