Luis Omar Becerra and Salvatore Lorefice 2009 Metrologia 46 07006 doi:10.1088/0026-1394/46/1A/07006
Luis Omar Becerra1 and Salvatore Lorefice2
Show affiliationsHydrometers are instruments usually made of glass which are widely used for different levels of precision to measure liquid density and related quantities to control different products and processes. This bilateral comparison on the calibration of hydrometers shows that results reported by CENAM–Mexico and INRIM–Italy are consistent within the claimed uncertainty in the range of 800 kg/m3 to 1200 kg/m3.
This bilateral comparison is intended to link the two regional comparisons SIM.M.D-K4 and EURAMET.D-K4.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by SIM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Issue 1A (Technical Supplement 2009)
Luis Omar Becerra and Salvatore Lorefice 2009 Metrologia 46 07006