C Kessler et al 2009 Metrologia 46 06012 doi:10.1088/0026-1394/46/1A/06012
C Kessler1, P J Allisy-Roberts1, D T Burns1, J Cardoso2 and C Oliveira2
Show affiliationsA first comparison of the standards for air kerma of the Instituto Tecnológico e Nuclear (ITN), Portugal and of the Bureau International des Poids et Mesures (BIPM) has been carried out in 137Cs gamma radiation. The comparison result, expressed as a ratio of the ITN standard and the recently re-evaluated BIPM standard, is 1.0013 (0.0021).
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).