Quick search Find article
Quick search
Find article

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60Co γ-rays

C Kessler1, P J Allisy-Roberts1, D T Burns1, P Roger1, L A de Prez2, J A de Pooter2 and P M G Damen2

Show affiliations


KEY COMPARISON

An indirect comparison of the standards for the quantity absorbed dose to water was carried out between the Van Swinden Laboratorium (VSL) and the Bureau International des Poids et Mesures (BIPM) in September 2005. The comparison was based on the calibration coefficients determined for three NE2611A type ionization chambers under reference conditions in the 60Co beams of both institutes. The comparison result, reported as a ratio of the VSL and the BIPM evaluations, is 0.9926 with a relative standard uncertainty of 4.9 × 10−3. This result replaces the previous indirect comparison result obtained in 2000 for the ongoing BIPM.RI(I)-K4 comparison. The degrees of equivalence for the VSL standard and the nine other national metrology institutes in this comparison are presented in the form of a matrix, together with those of a linked regional comparison. The opportunity was also taken to undertake a pilot study using the VSL water calorimeter at the BIPM.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A (Technical Supplement 2009)



  1. Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60Co γ-rays

    C Kessler et al 2009 Metrologia 46 06009

  2. Final report on subsequent key comparison CCQM-K18.1: pH of carbonate buffer

    L Vyskocil et al 2008 Metrologia 45 08014

  3. APMP/TCRI key comparison report of measurement of air kerma for medium-energy x-rays (APMP.RI(I)-K3)

    J H Lee et al 2008 Metrologia 45 06012

  4. Final report on the bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the CMI (Czech Republic) and the BIPM

    R Goebel et al 2008 Metrologia 45 01004

  5. Update of the ongoing comparison BIPM.RI(II)-K1.Ga-67 to include the activity measurements of the LNE–LNHB, France

    G Ratel et al 2006 Metrologia 43 06008

  6. An intercomparison of low flow gas facilities at eleven European laboratories using a Molbloc transfer package: EURAMET.MM.FF-S3 (EURAMET Project no 806)

    C Probert et al 2008 Metrologia 45 07012

  7. Final report of the on-going key comparison BIPM.QM-K1: Ozone at ambient level, comparison with CHMI, 2007

    Joële Viallon et al 2008 Metrologia 45 08005

  8. Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NIM and the BIPM in medium-energy x-rays

    D T Burns et al 2007 Metrologia 44 06008

  9. Activity measurements of the radionuclide 201Tl for the LNE–LNHB, France, PTB, Germany and the NPL, UK in the ongoing comparison BIPM.RI(II)-K1.Tl-201

    G Ratel et al 2008 Metrologia 45 06008

  10. Design of a 450 MHz β = 0.2 single spoke cavity at PKU

    Yao Zhong-Yuan et al 2009 Chinese Phys. C 33 292

View by subject




Export





Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.