Final report on APMP.PR-S1.1: Bilateral comparison of irradiance responsivity of UVA detectors

Xuebo Huang

Show affiliations


SUPPLEMENTARY COMPARISON

In order to assess the performance of the standards and techniques used for calibration and measurement of UVA irradiance responsivity of photodetectors in NMISA, South Africa, a new comparison was decided as a follow-up to comparison APMP.PR-S1. It is registered in the Key Comparison Data Base (KCDB) of BIPM as a bilateral supplementary comparison, with the identifier APMP.PR-S1.1. The comparison was carried out following the same technical protocol as that of supplementary comparison APMP PR-S1.

The principle, organization and method of the comparison, as well as the preliminary measurements at the pilot laboratory NMC-A*STAR Singapore, were described in the Final Report of the APMP.PR-S1 comparison. The results of this bilateral comparison show that the NMISA's results lie within ±2% against the comparison reference values of APMP.PR-S1, which is a great improvement.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the APMP, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).