R Goebel et al 2009 Metrologia 46 01013 doi:10.1088/0026-1394/46/1A/01013
R Goebel1, O Power2, N Fletcher1 and M Stock1
Show affiliationsThis report describes the results obtained from a NML(Ireland)–BIPM bilateral comparison of 10 kΩ resistance standards in 2008. The comparison was carried out in the framework of the BIPM ongoing key comparison BIPM.EM-K13.b. Two BIPM 10 kΩ travelling standards of SR104 type were calibrated first at the BIPM, then at the NMLI and again at the BIPM after their return. The stability of the transfer standards was such that the uncertainty associated with the transfer was smaller than the uncertainty arising from the calibrations. The NMLI and the BIPM calibrations were found to be in good agreement, with a mean difference smaller than the expanded uncertainty (k = 2) of the comparison.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Issue 1A (Technical Supplement 2009)
R Goebel et al 2009 Metrologia 46 01013
S Solve et al 2009 Metrologia 46 01002
S Solve et al 2009 Metrologia 46 01010
R Goebel et al 2008 Metrologia 45 01001
R Goebel et al 2009 Metrologia 46 01011
A Tonina et al 2010 Metrologia 47 01002
O Power et al 2009 Metrologia 46 01008
S Solve et al 2008 Metrologia 45 01006
S Solve et al 2010 Metrologia 47 01004
R Goebel et al 2010 Metrologia 47 01005