SUPPLEMENTARY COMPARISON

The electrical conductivity of non-ferrous metals with conductivities in the range 2.2 MS/m to 35.8 MS/m, provided by NPL in the form of bars (600 mm by 80 mm by 10 mm) and square blocks (80 mm side by 10 mm), was measured using an AC method at NPL and a DC method at PTB and VSL. The purpose of the comparison was to establish if there is a difference in the electrical conductivity measured for AC and DC conditions. Conductivity standards are used by industries involved with non-destructive testing (NDT) and evaluation (NDE) to calibrate the instruments used. These instruments operate at frequencies of at least 60 kHz and it was therefore necessary to establish agreement between AC and DC methods.

For all cases the uncertainty for AC conductivity measurements is considerably larger than that for DC measurements due to the increased complexity of the unique NPL bridge method. For DC measurements on blocks the PTB and VSL values agreed within the measurement uncertainties provided. A small difference between the values for a bar made from Nordic gold was removed when the CMC uncertainty for PTB was used. A better agreement between the DC measurement for blocks is believed to be due to a reduced sensitivity to material homogeneity as well as fewer dimensional measurements for the Van der Pauw technique used. A significant difference between the measured AC and DC conductivities for the titanium bar was found. This is believed to be due to a variation in the properties of the material through the thickness. The AC method is surface sensitive and so such a change in properties will influence the value compared to DC methods that are a volume average.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by EUROMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A (Technical Supplement 2009)



  1. Final report on EUROMET Project 427, EUROMET.EM-S7: Comparison of AC and DC conductivity standards

    Stuart Harmon et al 2009 Metrologia 46 01009

  2. Gauged U(1)R symmetries and Fayet-Iliopoulos terms in 5D orbifold supergravity

    Hiroyuki Abe and Kiwoon Choi JHEP12(2004)069

  3. Fractional branes and wrapped branes

    Duiliu-Emanuel Diaconescu et al JHEP02(1998)013

  4. Report of the spectral irradiance comparison EURAMET.PR-K1.a.1 between MIKES (Finland) and NIMT (Thailand)

    M Ojanen et al 2009 Metrologia 46 02001

  5. CCQM-K11: The determination of glucose in serum

    Michael J Welch et al 2003 Metrologia 40 08003

  6. B → η(') form factors in QCD

    Patricia Ball and Gareth W. Jones JHEP08(2007)025

  7. Report on Key Comparison EUROMET.EM-K8.1 (DC voltage ratio)

    G Marullo Reedtz and R Cerri 2005 Metrologia 42 01003

  8. BIPM comparison BIPM.RI(II)-K1.F-18 of activity measurements of the radionuclide 18F

    G Ratel and C Michotte 2003 Metrologia 40 06005

  9. Numerical analysis of magneto-electrospinning

    L Xu and Y Wu 2008 J. Phys.: Conf. Ser. 96 012208

  10. Tubular solutions in NS5-brane, Dp-brane and macroscopic strings backgrounds

    Wung-Hong Huang JHEP02(2005)061

View by subject




Export





Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.