Quick search Find article
Quick search
Find article

CCQM-K32 key comparison and P84 pilot study: Amount of silicon oxide as a thickness of SiO2 on Si

M P Seah

Show affiliations


KEY COMPARISON

CCQM-K32 and P84 were conducted following the pilot study P-38 to demonstrate and document the capability of interested National Metrology Institutes to measure the amount of silicon oxide on silicon wafers expressed as a thickness of SiO2 for nominal thicknesses in the range 1.5 nm to 8 nm. 'Amount of substance' may be expressed in many ways and here the measurand is the thickness of the silicon oxide layer on each of a total of 9 samples of nominal thicknesses in the range 1.5 to 8 nm on (100) and (111) Si substrates, expressed as the thickness of SiO2. This report presents the results from K32 and P84. It includes the data received for the measured values and their associated uncertainties, at 95% confidence, for the 9 samples prior to the deadline for receipt of data. The materials are grown by thermal oxidation in very clean furnaces designed for high quality gate oxides on Si wafers in European and US facilities at the same time as those for the pilot study, P-38. Separate samples were provided to each institute in special containers limiting the carbonaceous contamination to below about 0.3 nm. The 9 samples included 5 samples of ultra-thin SiO2 on (100) orientated wafers of Si and 4 samples of ultra-thin SiO2 on (111) orientated wafers of Si.

The measurements from the 11 participating laboratories were conducted using ellipsometry, neutron reflectivity (NR), x-ray photoelectron spectroscopy (XPS) or x-ray reflectivity measurements (XRR), guided by the protocol developed in the pilot study P-38 and reproduced in the Appendix. The measurements are given in tables 2 and 3. A very small correction is then made for the different samples that each laboratory received as in table 4. Where appropriate, method offset values deduced from the pilot study P-38 are given in table 5 leading to comparative data in tables 6 and 7. Values for the key comparison reference values (KCRVs) and their associated uncertainties are made from the weighted means and the expanded weighted standard deviations of the means from table 6. This is provided in table 8. Graphical plots of equivalence from tables 6 and 8 are provided in figure 1 and equivalence statements are presented in Annex A. Additional XPS and XRR data from NMIJ for K32 were withdrawn from the KCRV evaluation and are given in Annex B.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A ( 1 January 2008)



  1. CCQM-K32 key comparison and P84 pilot study: Amount of silicon oxide as a thickness of SiO2 on Si

    M P Seah 2008 Metrologia 45 08013

  2. Final report on supplementary comparison EUROMET.M.P-S3 of absolute pressure standards in the barometric range from 80 kPa to 110 kPa

    Pierre Otal et al 2008 Metrologia 45 07004

  3. BIPM comparison BIPM.RI(II)-K1.Yb-169 of activity measurements of the radionuclide 169Yb and links for the 1997 regional comparison EUROMET.RI(II)-K2.Yb-169

    G Ratel et al 2004 Metrologia 41 06017

  4. CCQM-K28: Tributyltin in sediment

    Céline S J Wolff Briche et al 2006 Metrologia 43 08001

  5. Activity measurements of the radionuclide 124Sb by the LNE–LNHB, France as a pilot study for the ongoing comparison BIPM.RI(II)-K1.Sb-124

    C Michotte et al 2006 Metrologia 43 06007

  6. Results from the CI-2005 campaign at the BIPM of the BIPM.L-K11 ongoing key comparison

    L Robertsson et al 2005 Metrologia 42 04004

  7. Final report on EUROMET.T-K4 (EUROMET Project 820): Comparison of the realizations of the ITS-90 at the freezing points of Al (660.323 °C) and Ag (961.78 °C)

    D Heyer et al 2008 Metrologia 45 03003

  8. Final report on the fixed-point comparison of Au/Pt thermocouples (APMP supplementary comparison APMP.T-S4)

    Yong-Gyoo Kim et al 2008 Metrologia 45 03004

  9. Results from the CII-2004 campaign at the BIPM of the BIPM.L-K11 ongoing key comparison

    L Robertsson et al 2005 Metrologia 42 04003

  10. Update of the BIPM comparison BIPM.RI(II)-K1.Cr-51 to include new activity measurements for the LNE–LNHB (France) and the pilot study result of the LNMRI (Brazil)

    G Ratel et al 2009 Metrologia 46 06001

View by subject




Export







Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.