W Sabuga and J Forastieri 2008 Metrologia 45 07020 doi:10.1088/0026-1394/45/1A/07020
W Sabuga1 and J Forastieri2
Show affiliationsThis report presents results of a bilateral comparison between PTB and INTI in the range 1.5 kPa to 300 kPa of absolute gas pressure. The comparison was carried out in April–May 2008. The pilot laboratory was PTB. The pressure standards of the participants were pressure balances of the same type. The transfer standard was an electronic pressure gauge. Deviations of the transfer standard's pressures from the reference pressures generated by the laboratories' standards at specified pressures were reported by the participants and used for calculating degrees of equivalence between the laboratories. The results of the comparison presented in the Draft A report have induced INTI to revise their uncertainty by an increase in the uncertainty of the residual pressure. After this revision, the results presented in the actual report demonstrate equivalence within the expanded uncertainties (k = 2) reported by the laboratories at all pressures of the comparison. The degrees of equivalence stated in this comparison will provide a supporting basis for a new pressure CMC table of INTI.
Main text. To reach the main text of this Paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by EURAMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Issue 1A ( 1 January 2008)
W Sabuga and J Forastieri 2008 Metrologia 45 07020
Ofer Aharony et al JHEP06(2000)026
Tom Banks et al JHEP01(2001)031
Niklas Beisert et al J. Stat. Mech. (2007) P01021
H George et al 2009 New J. Phys. 11 113028
Dietrich Dopheide 2006 Metrologia 43 07001
Puxun Wu and Hongwei Yu JCAP10(2007)014
Tomas Janssen and Tomislav Prokopec JCAP05(2007)010
G Beutier et al 2009 New J. Phys. 11 113026
Djamel Allal 2009 Metrologia 46 01007