C Probert et al 2008 Metrologia 45 07012 doi:10.1088/0026-1394/45/1A/07012
C Probert1, A I Johns1 and Z Metaxiotou2
Show affiliationsThis report describes the results from the intercomparison of low flow gas facilities at eleven European laboratories. The organizations taking part were TUV NEL (UK), EIM (Greece), NMi (The Netherlands), CMI (The Czech Republic), INRIM (Italy), MIKES (Finland), METAS (Switzerland), PTB (Germany), LNE (France), FI (Denmark) and UME (Turkey). EIM acted as joint pilot laboratory with TUVNEL and NMi acted as assisting laboratory.
The transfer package consisted of two high-accuracy laminar-flow elements with associated hardware and software. The laminar-flow elements were Molblocs manufactured by DH Instruments Inc. The two Molblocs covered different flow ranges and were calibrated separately. The flow ranges covered were 2.1 mg/s to 21 mg/s for Molbloc A and 83 mg/s to 578 mg/s for Molbloc B.
Main text. To reach the main text of this Paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by EURAMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Issue 1A ( 1 January 2008)
C Probert et al 2008 Metrologia 45 07012