Final report on supplementary comparison APMP.M.P-S2 (bilateral comparison) at a nominal pressure of 0.05 Pa

Pardeep Mohan1 and Patrick J Abbott2

Show affiliations


SUPPLEMENTARY COMPARISON

A supplementary bilateral comparison was carried out to compare two different pressure standards at a nominal pressure of 0.05 Pa. The two standards used different principles for generating the nominal pressure; one based on the static expansion method (NPL, India), and the other based on the orifice flow or dynamic expansion method (NIST, USA). The transfer standard consisted of a pair of spinning rotor gauges whose accommodation coefficients were measured at a nominal pressure of 0.05 Pa at the two laboratories. The nominal pressure value for the comparison was chosen because (i) it can be easily generated by the respective standards with high accuracy and (ii) it had not been covered in the earlier two key comparisons, namely, CCM.P-K3 and CCM.P-K4, which the two NMIs had participated in. NPLI served as the pilot lab, and provided the stainless steel rotors; each lab used its own thimbles and electronics. The rotors were hand-carried from NPLI to NIST at atmospheric pressure in small, insulation-packed glass vials to prevent the rotors from moving during transit; otherwise no special transit precautions were taken.

The pressure value of 0.05 Pa generated by the NPLI and NIST showed good agreement within their combined uncertainties, proving thereby that the two standards were equivalent to each other for the pressure of 0.05 Pa.

Main text. To reach the main text of this Paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the APMP, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A ( 1 January 2008)



  1. Final report on supplementary comparison APMP.M.P-S2 (bilateral comparison) at a nominal pressure of 0.05 Pa

    Pardeep Mohan and Patrick J Abbott 2008 Metrologia 45 07003

  2. Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60Co γ-rays

    C Kessler et al 2009 Metrologia 46 06009

  3. Final report on subsequent key comparison CCQM-K18.1: pH of carbonate buffer

    L Vyskocil et al 2008 Metrologia 45 08014

  4. APMP/TCRI key comparison report of measurement of air kerma for medium-energy x-rays (APMP.RI(I)-K3)

    J H Lee et al 2008 Metrologia 45 06012

  5. Final report on the bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the CMI (Czech Republic) and the BIPM

    R Goebel et al 2008 Metrologia 45 01004

  6. Update of the ongoing comparison BIPM.RI(II)-K1.Ga-67 to include the activity measurements of the LNE–LNHB, France

    G Ratel et al 2006 Metrologia 43 06008

  7. An intercomparison of low flow gas facilities at eleven European laboratories using a Molbloc transfer package: EURAMET.MM.FF-S3 (EURAMET Project no 806)

    C Probert et al 2008 Metrologia 45 07012

  8. Final report of the on-going key comparison BIPM.QM-K1: Ozone at ambient level, comparison with CHMI, 2007

    Joële Viallon et al 2008 Metrologia 45 08005

  9. Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NIM and the BIPM in medium-energy x-rays

    D T Burns et al 2007 Metrologia 44 06008

  10. Activity measurements of the radionuclide 201Tl for the LNE–LNHB, France, PTB, Germany and the NPL, UK in the ongoing comparison BIPM.RI(II)-K1.Tl-201

    G Ratel et al 2008 Metrologia 45 06008

View by subject




Export







Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.