Final report on CCM key comparison CCM.P-K2: Pressure (10 kPa to 120 kPa) absolute mode

Michael Perkin1, Alain Picard2, Michel Lecollinet3, Kitty Fen4, Marina Sardi5, Archie Miiller6, Anil Agarwal7, Michael Jescheck8 and Christian Wüthrich9

Show affiliations


KEY COMPARISON

The results are presented of key comparison CCM.P-K2 that covered the pressure range 10 kPa to 120 kPa in absolute mode. Nine national measurement institutes participated in the comparison, which used a piston–cylinder assembly of 335 mm2 nominal effective area as a transfer standard.

Whilst not all laboratories were found to be equivalent at all pressures, the larger-than-expected instabilities in the transfer standard precluded the ability to discriminate between participants' standards at a level needed to fully support the Calibration Measurement Capabilities table in the BIPM key comparison database.

Main text. To reach the main text of this Paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A ( 1 January 2008)



  1. Final report on CCM key comparison CCM.P-K2: Pressure (10 kPa to 120 kPa) absolute mode

    Michael Perkin et al 2008 Metrologia 45 07002

  2. Update of the BIPM comparison BIPM.RI(II)-K1.Ho-166m activity measurements to include the IRA and the NPL and a re-evaluation of the degrees of equivalence for the APMP.RI(II)-K2.Ho-166m comparison

    C Michotte et al 2009 Metrologia 46 06008

  3. EUROMET project no 429: Comparison of 10 V electronic voltage standards

    F Liefrink et al 2003 Metrologia 40 01004

  4. Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kV to 300 kV

    L Büermann et al 2008 Metrologia 45 06013

  5. Final report on EUROMET supplementary comparison EUROMET.L-S12: Calibration of gauge blocks by mechanical comparison

    Howard McQuoid and Zbigniew Ramotowski 2008 Metrologia 45 04002

  6. Magnons, their solitonic avatars and the Pohlmeyer reduction

    Timothy J. Hollowood and J. Luis Miramontes JHEP04(2009)060

  7. Update of the BIPM comparison BIPM.RI(II)-K1.F-18 of activity measurements of the radionuclide 18F to include the PTB

    G Ratel et al 2006 Metrologia 43 06001

  8. Results from the CI-2004 campaign at the BIPM of the BIPM.L-K11 ongoing key comparison

    L Robertsson et al 2005 Metrologia 42 04002

  9. International Key Comparison CCQM-K26.b and Pilot Study CCQM-P50.b: Comparison of primary standards of sulphur dioxide (SO2) in synthetic air

    Masaaki Maruyama et al 2007 Metrologia 44 08007

  10. Final report on comparison COOMET.T-K3: Regional comparisons of the national standards of temperature in the range from the triple point of water to the freezing temperature of zinc

    A I Pokhodun and A G Ivanova 2008 Metrologia 45 03005

View by subject




Export





Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.