C K Ross et al 2008 Metrologia 45 06010 doi:10.1088/0026-1394/45/1A/06010
C K Ross1, K R Shortt1, M Saravi2, A Meghzifene3, V M Tovar4, R A Barbosa5, C N da Silva5, L Carrizales6 and S M Seltzer7
Show affiliationsTransfer chambers were used to compare the standards for 60Co air kerma maintained by seven laboratories. Six of the laboratories are members of the Sistema Interamericano de Metrología (SIM) regional metrology organization while the seventh is the International Atomic Energy Agency (IAEA) laboratory in Vienna. The National Research Council (NRC) acted as the pilot laboratory for the comparison. Because of the participation of laboratories holding primary standards, the comparison results could be linked to the key comparison reference value maintained by the Bureau International des Poids et Mesures (BIPM). The results for all laboratories were within the expanded uncertainty (two standard deviations) of the reference value. The estimated relative standard uncertainty of the comparison between any pair of laboratories ranged from 0.5% to 1.0%. The largest discrepancy between any two laboratories was 1.0%.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Issue 1A ( 1 January 2008)
C K Ross et al 2008 Metrologia 45 06010
Robert R Greenberg 2008 Metrologia 45 08016
A K Bandyopadhyay et al 2008 Metrologia 45 07001
R Goebel and M Stock 2004 Metrologia 41 02004
S Solve et al 2008 Metrologia 45 01006
G Ratel et al 2006 Metrologia 43 06002
N Yamada et al 2008 Metrologia 45 04001
N G Domostroeva and A G Chunovkina 2008 Metrologia 45 07009
G Ratel et al 2007 Metrologia 44 06002
G Ratel et al 2008 Metrologia 45 06009