Final report of the SIM 60Co air-kerma comparison SIM.RI(I)-K1

C K Ross1, K R Shortt1, M Saravi2, A Meghzifene3, V M Tovar4, R A Barbosa5, C N da Silva5, L Carrizales6 and S M Seltzer7

Show affiliations


KEY COMPARISON

Transfer chambers were used to compare the standards for 60Co air kerma maintained by seven laboratories. Six of the laboratories are members of the Sistema Interamericano de Metrología (SIM) regional metrology organization while the seventh is the International Atomic Energy Agency (IAEA) laboratory in Vienna. The National Research Council (NRC) acted as the pilot laboratory for the comparison. Because of the participation of laboratories holding primary standards, the comparison results could be linked to the key comparison reference value maintained by the Bureau International des Poids et Mesures (BIPM). The results for all laboratories were within the expanded uncertainty (two standard deviations) of the reference value. The estimated relative standard uncertainty of the comparison between any pair of laboratories ranged from 0.5% to 1.0%. The largest discrepancy between any two laboratories was 1.0%.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A ( 1 January 2008)



View by subject




Export







Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.