Final report on the thermal expansion coefficient of gauge blocks (APMP.L-S1)

N Yamada1, H Matsumoto1, M Sassi2, R Thalmann3, A Lassila4, R Schoedel5, H A Castollo6 and P Balling7

Show affiliations


SUPPLEMENTARY COMPARISON

This report describes the results obtained from a supplementary comparison on the thermal expansion coefficient of gauge blocks in 2004–2006. The comparison was carried out by members in DG8 of CCL. The four gauge blocks, which were three ceramics blocks and one steel block, were calibrated by seven organizations in turn. Most reported results corresponded with each other within their expanded uncertainties (k = 2). However, some results for the steel gauge block had a systematic deviation from the key reference value. It was considered that the source of the deviation was mainly caused by the capability of the temperature determination of the gauge block.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the APMP, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A ( 1 January 2008)



  1. Final report on the thermal expansion coefficient of gauge blocks (APMP.L-S1)

    N Yamada et al 2008 Metrologia 45 04001

  2. COOMET.M.V-K1 key intercomparison of liquid viscosity measurements

    N G Domostroeva and A G Chunovkina 2008 Metrologia 45 07009

  3. BIPM comparison BIPM.RI(II)-K1.Np-237 of activity measurements of the radionuclide 237Np and links for the 1998 regional comparison EUROMET.RI(II)-K2.Np-237

    G Ratel et al 2007 Metrologia 44 06002

  4. Activity measurements of the radionuclide 54Mn for the LNE–LNHB, France, NPL, United Kingdom and the MKEH, Hungary, in the ongoing comparison BIPM.RI(II)-K1.Mn-54

    G Ratel et al 2008 Metrologia 45 06009

  5. Final report on intercomparison measurements in the field of gas flow and volume, which are realized at two levels: the second level (B*)

    Stefan Makovnik 2008 Metrologia 45 07018

  6. Final report of the on-going key comparison BIPM.QM-K1: Ozone at ambient level, comparison with NIST, 2007

    Joële Viallon et al 2008 Metrologia 45 08008

  7. Modeling and characterization of electrostatic comb-drive actuators in conducting liquid media

    Vikram Mukundan et al 2009 J. Micromech. Microeng. 19 065008

  8. Comparison of the standards for absorbed dose to water of the PTB, Germany and the BIPM for 60Co γ rays

    C Kessler et al 2006 Metrologia 43 06005

  9. SIM international comparison of 50/60 Hz energy (2002–2007)

    Tom Nelson et al 2008 Metrologia 45 01009

  10. Final report on the bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the NIST (USA) and the BIPM

    R Goebel et al 2008 Metrologia 45 01001

View by subject




Export







Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.