Final report on EUROMET.T-K4 (EUROMET Project 820): Comparison of the realizations of the ITS-90 at the freezing points of Al (660.323 °C) and Ag (961.78 °C)

D Heyer1, U Noatsch1, E Tegeler1, M Anagnostou2, E Turzo-Andras3, I Antonsen4, V Augevicius5, J Bojkovski6, A Bronnum7, V Chimenti8, S Duris9, E Filipe10, S Gaita11, J Gray12, D Head12, E Grudniewicz13, J Ivarsson14, M Kalemci15, O Kerkhof16, I Lobo10, S Nemeth3, A Pokhodun17, J Ranostaj9, E Renaot18, P Rosenkranz19, M Smid20, P Steur21, A Steiner22, M Valin18, T Veliki23 and T Weckström24

Show affiliations


KEY COMPARISON

The comparison EUROMET.T-K4 is the regional extension of CCT-K4: an intercomparison of the realisations of the freezing points of Al (660.323 °C) and Ag (961.78 °C). The intercomparison was organized in four loops with long stem standard platinum resistance thermometers (SPRTs) as travelling standards: One 25 Ω thermometer to be used only at the Al freezing point and two high temperature SPRTs to be used only at the Ag freezing point in each loop. Parallel to the measurements with thermometers the pilot and sub-pilots organized an internal intercomparison using an Ag fixed point cell.

Most HTSPRTs showed a strong drift which is mainly due to mechanical stress and poisoning of the sensor by impurities. This drift can partly be compensated by a correction applying Matthiessen's rule. An evaluation of the data taking into account both HTSPRTs in each loop, the linkage of the sub-pilots by measurements at the Ag freezing point and a possible compensation according to Matthiessens's rule allows to calculate the results of the participants also for the measurements at the Ag freezing point. European reference values (ERV) were calculated for measurements at the Al and Ag freezing points. The difference between these ERVs and the corresponding KCRVs of CCT-K4 is also given, which allows the linking of the different comparisons.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).


Dates

Issue 1A ( 1 January 2008)



  1. Final report on EUROMET.T-K4 (EUROMET Project 820): Comparison of the realizations of the ITS-90 at the freezing points of Al (660.323 °C) and Ag (961.78 °C)

    D Heyer et al 2008 Metrologia 45 03003

  2. Final report on the fixed-point comparison of Au/Pt thermocouples (APMP supplementary comparison APMP.T-S4)

    Yong-Gyoo Kim et al 2008 Metrologia 45 03004

  3. Results from the CII-2004 campaign at the BIPM of the BIPM.L-K11 ongoing key comparison

    L Robertsson et al 2005 Metrologia 42 04003

  4. Update of the BIPM comparison BIPM.RI(II)-K1.Cr-51 to include new activity measurements for the LNE–LNHB (France) and the pilot study result of the LNMRI (Brazil)

    G Ratel et al 2009 Metrologia 46 06001

  5. Link between the comparison EUROMET.EM.BIPM-K11.b and the ongoing comparison BIPM.EM-K11.b

    Dominique Reymann 2003 Metrologia 40 01005

  6. Final report on supplementary comparison APMP.M.P-S2 (bilateral comparison) at a nominal pressure of 0.05 Pa

    Pardeep Mohan and Patrick J Abbott 2008 Metrologia 45 07003

  7. Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60Co γ-rays

    C Kessler et al 2009 Metrologia 46 06009

  8. Final report on subsequent key comparison CCQM-K18.1: pH of carbonate buffer

    L Vyskocil et al 2008 Metrologia 45 08014

  9. APMP/TCRI key comparison report of measurement of air kerma for medium-energy x-rays (APMP.RI(I)-K3)

    J H Lee et al 2008 Metrologia 45 06012

  10. Final report on the bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the CMI (Czech Republic) and the BIPM

    R Goebel et al 2008 Metrologia 45 01004

View by subject




Export





Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.