Bilateral direct comparison of Josephson array voltage standards of the PTB (Germany) and PEL (Croatia): Draft B report for EUROMET.EM-S28

Ralf Behr1, Damir Ilić2 and Alan Šala2

Show affiliations


SUPPLEMENTARY COMPARISON

A comparison of the 10 V Josephson array voltage standard of the Physikalisch-Technische Bundesanstalt (PTB), Germany, was made with that of the Primarni Elektromagnetski Laboratorij (PEL), Croatia, in April 2007. The results are in very good agreement and the measured overall relative combined standard uncertainty is better than 1 part in 1010.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by EURAMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).