R Goebel et al 2008 Metrologia 45 01004 doi:10.1088/0026-1394/45/1A/01004
R Goebel1, P Chrobok2, N Fletcher1 and M Stock1
Show affiliationsThis report describes the results obtained from a CMI–BIPM bilateral comparison of 1 Ω resistance standards in 2007–2008. The comparison was carried out in the framework of the BIPM ongoing key comparison BIPM.EM-K13.a. Two BIPM 1 Ω travelling standards of CSIRO type were calibrated first at the BIPM, then at the CMI and again at the BIPM after their return. The stability of the transfer standard was such that the uncertainty associated with the transfer was significantly smaller than the uncertainty arising from the calibrations. The CMI and the BIPM calibrations were found to be in good agreement, with a mean difference smaller than the expanded uncertainty (k = 2) of the comparison.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).