O Power and N Fletcher 2008 Metrologia 45 01002 doi:10.1088/0026-1394/45/1A/01002
O Power1 and N Fletcher2
Show affiliationsA comparison of 100 pF capacitance standards at a measuring frequency of 1 kHz was carried out between the BIPM and the NML (Ireland). Two travelling standards belonging to the BIPM were used. Measurements were made over the period June to October 2007. One of the travelling standards had to be eliminated from the comparison due to a failure of its temperature control circuit. The transport characteristics of the remaining standard were satisfactory, and the results showed agreement well within the combined uncertainty of the comparison (relative expanded uncertainty 0.78 ppm, k = 2, 95% confidence).
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).