I A Robinson and B P Kibble 2007 Metrologia 44 427 doi:10.1088/0026-1394/44/6/001
I A Robinson and B P Kibble
Show affiliationsDuring 2006 modifications were made to the NPL Mark II moving-coil watt balance which eliminated several significant sources of error. The apparatus was run from October 2006 to March 2007 to check its operation and to produce a preliminary value of Planck's constant h. This paper reports the current state of the work and predicts the uncertainty that is expected to be achievable with the existing apparatus. The preliminary value of h is 6.626 070 95(44) × 10−34 J s, which corresponds to a value of the Avogadro constant of NA = 6.022 139 99(40) × 1023 mol−1.
07.50.-e Electrical and electronic instruments and components
Issue 6 (December 2007)
Received 10 July 2007
Published 11 October 2007
I A Robinson and B P Kibble 2007 Metrologia 44 427
John Ancsin 2008 Metrologia 45 16
K Nagasaki and N Yanagi 2002 Plasma Phys. Control. Fusion 44 409
Jing-yu Wang et al 2005 Semicond. Sci. Technol. 20 L36
Jun Ouyang et al 2005 Smart Mater. Struct. 14 524
D. Biskamp and E. Schwarz 1997 Europhys. Lett. 40 637
Fatima Anis et al 2009 J. Phys. B: At. Mol. Opt. Phys. 42 091001
Bishu Agarwal et al 2009 Phys. Scr. 80 055505
G Basile et al 1991 Metrologia 28 455
Anton M. Zeitlin JHEP12(2007)098