D T Burns et al 2007 Metrologia 44 06008 doi:10.1088/0026-1394/44/1A/06008
D T Burns1, Tian Zhongqing2, Lu Yazhu2, Wan Guoqing2 and Fan Cheng2
Show affiliationsA key comparison has been made between the air-kerma standards of the NIM and the BIPM in the medium-energy x-ray range. The results show the standards to be in general agreement at the level of twice the stated relative standard uncertainty of 3.1 × 10-3, although there is evidence of a trend in the results at different radiation qualities. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).