The triple point of nitrogen

L Lipiński1, A Kowal1, A Szmyrka-Grzebyk1, H Manuszkiewicz1, P P M Steur2 and F Pavese2

Show affiliations


We present accurate measurements of the triple point of nitrogen. The quality of this triple point is once more confirmed, and its temperature value is found to be T90 = 63.1526 K (expanded uncertainty U = 0.30 mK, reproducibility better than 0.1 mK). This is the first direct determination of this point on ITS-90 since the introduction of the scale. This point is useful as a non-uniqueness checkpoint.


Footnote
*  Formerly Istituto di Metrologia 'G Colonnetti'.
PACS

06.20.Dk Measurement and error theory

07.20.Dt Thermometers

06.20.fb Standards and calibration

07.20.Mc Cryogenics; refrigerators, low-temperature equipment

Subjects

Instrumentation and measurement

Dates

Issue 5 (October 2006)

Received 31 May 2005

Published 21 September 2006