Quick search Find article
Quick search
Find article

CCQM-P39.1: As, Hg, Pb, Se and methylmercury in salmon

Y Aregbe and P D P Taylor

Show affiliations


PILOT STUDY

CCQM-P39.1 was organized as a follow-up pilot study in parallel to the key comparison CCQM-K43 after the previous pilot study on tuna fish. CCQM-P39.1 was an activity of the Inorganic Analysis Working Group (IAWG) of CCQM and was coordinated by the Joint Research Centre–Institute for Reference Materials and Measurements (IRMM, Geel, Belgium) of the European Commission (EC). In CCQM-P39.1 the amount contents of As, Hg, Pb, Se and methylmercury (CH3Hg) in salmon (muscle and skin) were the measurands under investigation. Besides the national metrology institutes (NMIs) also non-IAWG members, expert laboratories for mercury and methylmercury measurements, were invited to participate in this pilot study. Results were reported by six IAWG members and six expert laboratories. During the CCQM-IAWG autumn meeting in Berlin, October 2005, it was agreed that in CCQM-K43 the KCRV is calculated as the mixture model median (MM-median) of all reported results. Therefore in CCQM-P39.1 the reported results are presented graphically with the KCRV from CCQM-K43.

The reported results of the IAWG members fall within a range of ±4% for arsenic and lead relative to the CCQM-K43 KCRV. For mercury, the spread was ±2%, but one IAWG member reported a very large uncertainty on the measurement result. For selenium the spread of IAGW members is ±2% deviation from the CCQM-K43 KCRV. Including the reported results from the invited expert laboratories, the spread of results increased for arsenic, lead and mercury to ±8%. The reported results including the experts fall within a range of ±20% for selenium and ±30% for methylmercury.

The methods applied were isotope dilution mass spectrometry (IDMS) using sector field or quadrupole inductively coupled plasma-mass spectrometry (ICP-MS), external calibration or standard addition using ICP-MS, atomic absorption spectrometry (AAS), atomic fluorescence spectrometry (AFS), atomic emission detection (AED) and electron capture detection (ECD). Instrumental neutron activation analysis (INAA) and k0-neutron activation analysis (k0-NAA) were also used as analytical techniques. Each of the five analytes under investigation was measured with at least two of these analytical techniques.

IAEA reported two results for arsenic and lead. They were measured with ICP-MS and AAS, respectively. JSI reported two results for mercury, measured with NAA and AAS, and furthermore two results for methylmercury measured with GC-ECD and GC-CV-AFS.

This report presents the participants' results in CCQM-P39.1 for all analytes under investigation. In Annex 1, the results are displayed with the CCQM-K43 KCRV. In Annex 2, the different approaches for methlymercury measurements are presented in more detail. In Annex 3, the questionnaire data are presented. Annex 4 compiles all the CCQM-P39.1 information documents.

The key comparison CCQM-K43 was carried out in parallel to this pilot study for the same measurands in the same salmon material. Participation was meant for IAWG members, to support their CMCs in Appendix C.

Main text. To reach the main text of this paper, click on Final Report.

The final report has been peer-reviewed and approved for publication by the CCQM Working Group on Inorganic Analysis.


Dates

Issue 1A ( 1 January 2006)



  1. CCQM-P39.1: As, Hg, Pb, Se and methylmercury in salmon

    Y Aregbe and P D P Taylor 2006 Metrologia 43 08006

  2. Non local theories: new rules for old diagrams

    Gherardo Piacitelli JHEP08(2004)031

  3. FOCUS ON ATTOSECOND PHYSICS

    André D Bandrauk et al 2008 New J. Phys. 10 025004

  4. Supergravity S-Branes

    Martin Kruczenski et al JHEP05(2002)039

  5. Magnetic impurities in the pnictide superconductor Ba1−xKxFe2As2

    Sutirtha Mukhopadhyay et al 2009 New J. Phys. 11 055002

  6. CCQM-P39: As, Hg, Pb, Se and methylmercury in tuna fish

    Y Aregbe et al 2004 Metrologia 41 08004

  7. CCQM-P32 pilot study. Anion calibration solutions

    Michael Weber and Jürg Wüthrich 2003 Metrologia 40 08007

  8. Environment-assisted quantum transport

    Patrick Rebentrost et al 2009 New J. Phys. 11 033003

  9. Correlated counting of single electrons in a nanowire double quantum dot

    Theodore Choi et al 2009 New J. Phys. 11 013005

  10. CCQM-P43: Tributyltin and dibutyltin in sediment

    Céline S J Wolff Briche et al 2006 Metrologia 43 08002

View by subject




Export







Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.