KEY COMPARISON

A bilateral comparison of illuminance responsivity scales between the UME and the IFA-CSIC was carried out, where the IFA-CSIC acted as the pilot and link to the key comparison CCPR-K3.b. The ratio of the measured illuminance responsivities (UME/IFA) was 1.0003 with expanded uncertainty of 0.0084 (k = 2) including the uncertainty of the comparison and the uncertainties of the realization of the scales.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCPR, according to the provisions of the Mutual Recognition Arrangement (MRA).