Ian Veldman and Hans-Jürgen von Martens 2004 Metrologia 41 09001 doi:10.1088/0026-1394/41/1A/09001
Ian Veldman1 and Hans-Jürgen von Martens2
Show affiliationsA supplementary comparison (SADC.AUV.V-S1) was organized to compare measurements of sinusoidal linear accelerations in the frequency range from 10 Hz to 10 kHz. In this bilateral comparison between the CSIR National Metrology Laboratory (CSIR–NML), South Africa and the Physikalisch-Technische Bundesanstalt (PTB), Germany both NMIs applied laser interferometry in compliance with method 3 of the international standard ISO 16063-11:1999.
The complex charge sensitivity (magnitude and phase) of two different transfer standards (single-ended accelerometers) was measured at 37 frequencies. The results of the SADCMET.AUV.V-S1 are a set of comparison values and their uncertainties, and degrees of equivalence between these and between the two laboratories. From this complete set of results, six matrices of equivalence per accelerometer were selected and are demonstrated graphically.
The deviations between the PTB and NML results were smaller than 0.6° and 2° for the phase shift measurements for the Endevco and Brüel & Kjær accelerometers respectively (37 measurement points each).
Main text. To reach the main text of this paper, click on Final Report.
The final report has been peer-reviewed and approved for publication by SADCMET, according to the provisions of the Mutual Recognition Arrangement (MRA).
Issue 1A ( 1 January 2004)
Ian Veldman and Hans-Jürgen von Martens 2004 Metrologia 41 09001
André-Claude Clapson et al 2008 Class. Quantum Grav. 25 035002
The HERMES collaboration JHEP11(2009)083
Ilka Brunner et al JHEP11(2006)006
Shunichi Wakitani 2007 J. Phys.: Conf. Ser. 64 012006
Frank Guenther and Alejandro Perez-Castorena 2009 Metrologia 46 08001
R G Singh et al 2009 J. Phys. D: Appl. Phys. 42 062002
Santanu K Maiti 2009 Phys. Scr. 80 055704
Ivan Andric and Danijel Jurman JHEP01(2005)039
Jung A Lee et al 2008 J. Micromech. Microeng. 18 035012